Hygiene and maintenance
Universal support
W&H recommends sterilization according to ANSI/AAMI ST55, ANSI/AAMI ST79.
> Read the notes, follow the instructions and heed the warnings provided by the manufacturers of steam sterilizers.
> The program selected must be suitable for the universal support.
Recommended sterilization procedures
> “Dynamic-air-removal cycle” (type B, S)
132°C (270°F) for at least 4 minutes
> “Gravity-displacement cycle” (type N)
121°C (250°F) for at least 30 minutes
> Maximum sterilization temperature 135°C (275°F)
Sterilization