7.2.2 Primary Data
The insert in the Primary Data window display the fluorescence
trace during and right after the last saturation pulse. In addition,
the window repeats the data of the previous one (Ft, F
V
/F
M
or
Y(II), ETR, PAR). New information of the Primary Data window is
the F
0
and F
M
signal levels (after Fo,Fm was pressed) or the F
and F
M
’ signal levels (after SAT was pressed). Additionally,
Primary Data
A
SAT
MENU
Act.L.
Side Panel
Same as Basic Data win-
dow (Fig. 27, page 53) plus
upward arrow key.
Ft
583
F
1289
Fm’
1590
Y(II)
0.184
ETR
32.5
PAR
420 O2
-
Temp
-
Aux1
-
Hum
-
Aux2
-
Rec Fo,Fm Clock
Mark
2018-04-12
8:20:09
Main Panel
Ft
, current fluorescence level
Displayed after Fo,Fm :
Fo
,
Fm
, and
Fv/Fm
; minimum and maximum fluores-
cence levels of dark-acclimated sample,
maximum photochemical PS II yield. Dis-
played after SAT :
F
,
Fm’
, and
Y(II)
; fluo-
rescence level just before saturation pulse
and maximum fluorescence of light-
exposed sample, effective photochemical
quantum yield of PS II
Insert
, fluorescence trace during last
saturation pulse.
PAR,
Temp
, and
Hum
, photosynthet-
ic active radiation in µmol·m
-2
·s
-1
and
temperature in °C, respectively.
O2
,
Aux1
,
Aux2
, data measured by
optional oxygen, humidity and other
sensors.
Rec , Fo,Fm , Clock , Mark : see
Fig. 28: