26
and surface finish. If not set manually, the system will use the default
value.
Appendix A
provides the emissivity value of common materials.
Customize:
Available on 50 Series Inspectors only
TCorrect:
Available on 50 Series Inspectors only
TReflect:
When Set to
On
, allows user to enter value of reflected
temperatures that may be in viewed image.
4.4.3
Alarms
Press
Confirm
key to activate the main menu. Select Setup->Alarms
and press
Confirm
key to display the Analysis dialog box. User can set
temperature alarm, alarm temperature, alarm color, isotherm color, isotherm
temperature and isotherm width. Press
Up
and
Down
keys to choose menu
items,
Left
and
Right
keys to modify the item contents.
An alarm is triggered when a measurement cursor’s measurement meets or
exceeds the
Set Alarm Temp
setting. All pixels above this temperature will
change to the color as set in
Alarm Color
. When the
Alarm Color
is set to
None
, only the audible alarm will sound. When using alarms with
Area
measurements, the selected measurement mode,
Min
,
Max
or
Average
will
have to meet or exceed the set temperature in order to trigger the alarm.