C
HAPTER
4
L
IMIT
A
NALYSER
O
PERATING THE
S
YSTEM
P
AGE
29
4.3 Setting the program switches
The Limit Analyser is programmed using small dual-in-line-switch S1-S10 and
adjusted to specific operating requirements.
switch off the unit and open the lid of he controller
→ WARNING supply voltage 85 … 264 V AC 47 … 63Hz
→ The DIP-switches (dual in-line package) are located between
the LED lights and the push buttons. Next to the switches
you see the potentiometer for setting the flushing time.
O N
S P Ü L Z E IT
1
2
3
4
5
6
7
9
8
1 0
factory settings
flushing time
ca. 2 minutes
Poti app.
10:00
analysis interval
10 minutes
S1 OFF
S2 ON
retry cycle
yes
S3 ON
relay function
permanent
contact
without analysis stop
S4 ON
S5 OFF
parameter
total hardness
S6 OFF
S7 OFF
S8 OFF
operating mode
analysis mode
S9 OFF
S10 OFF
Work on electrical equipment of the plant / machinery must be performed
by a qualified electrician! Observe all applicable electrical installation
work rules.
→ WARNING supply voltage 85 … 264 V AC 47 … 63Hz
For the operation of the rotary potentiometer and for setting the slide
switch you need a small insulated screwdriver
→
Please use only good, proven tools, you will help avoid damage to
sensitive components
WARNING
ELECTRICITY
NOTE
F
LUSHING
T
IME
D
IP
S
WITCHES