C
HAPTER
5
S
YCON
2500
M
AINTAINANCE
P
AGE
41
M
AINTENANCE
S
TEP
3
FLUSH
INDICATOR
+
R
EFILL
I
NDICATOR AND
R
ESET THE
S
TORAGE
C
OUNTER
1. remove empty indicator bottle
→ open bottle cap by turning
→ remove the suction pipe carefully
→ If indicator fluid leaks, wipe up with a paper
towel
→ dispose of empty bottle
2. replace full indicator bottle
→ remove the cap
→ insert the suction pipe carefully
→ screw bottle cap by turning
When you are installing pay attention to a
correct position of the screw and the adapter.
3. de-aerate the indicator hose pipe
→ fill measuring chamber with water
→ press FLUSH button several times
→ dose indicator until the hoses and the
cartridge of peristaltic pump are completely
filled with indicator and no air bubbles are
visible
→ press INDICATOR button several times
4. confirm bottle change and reset counter
→ press INDICATOR and RESET button
simultaneously until the right red and the blue
LED light up briefly
blue
red
NOT