4.4
Start up in 5 steps
Make sure that the analyser was installed in accordance with chapter 3
(→ page 19) and the program switches were set properly according to the
application like described under Section 4.3
(→ page 29).
step 1
switch on the unit
→ switch power button „on“
→ green LED
"analysis result" flashes.
green
step 2
+
reset indicator storage volume
→ Press the INDICATOR and RESET
button at the same time
WARNING !!I
run reset function only when replace
empty bottle by a full one
(→ page 41)
step 3
flood measuring chamber with water
→ Press the FLUSH button until the
measuring chamber is completely filled
with water and free of bubbles.
step 4
de-aerate dosing pump
→ Press the INDICATOR button until
indicator is delivered continuously into
the measuring chamber. During this
process, the agitator rotates.
step 5
start analysis
→ press START button to run first
analysis
→ an analysis starts with the flushing of
the measuring chamber
START
INDICATO R
RESET
FLUSH
INDICATO R