![VWR MD 8000 L Operating Manual Download Page 75](http://html1.mh-extra.com/html/vwr/md-8000-l/md-8000-l_operating-manual_1057579075.webp)
MD 8000 L
Data storage
ba77218e01
11/2018
75
12.2 Automatic data storage at intervals
The storage interval (
Interval
) determines the time interval between automatic
data storage processes. Each data storage process transmits the current data-
set to the USB interface.
Configuring the
automatic memory
function
Settings
You can configure the automatic data storage function with the following set-
tings:
1.
Press the
<STR
_
>
key.
The menu for automatic data storage appears.
1 Specified entire storing duration
2 Max. available storage duration
3 Graphical display of the storage usage
ID number
1
Interval 30
s
Duration 180
min
continue
0d03h00min
0
1d17h33min
Automatic data storage
1
2
3
01.11.2018
08:00
Back
Menu item
Possible setting
Explanation
ID number
1 ... 10000
ID number for the dataset series.
Interval
1 s, 5 s, 10 s,
30 s, 1 min,
5 min, 10 min,
15 min, 30 min,
60 min
Storage interval.
The lower limit of the storage interval
can be restricted by the number of free
storage locations.
The upper limit is restricted by the stor-
age duration.
Duration
1 min ... x min
Storage duration.
Specifies after which time the auto-
matic data storage should be termi-
nated.
The lower limit of the storage duration
is restricted by the storage interval.
The upper limit is restricted by the
number of free storage locations.
Summary of Contents for MD 8000 L
Page 93: ...MD 8000 L Glossary ba77218e01 11 2018 93...
Page 97: ......