MD 8000 L
Settings
ba77218e01
11/2018
71
11.6.4 Automatic switch-off function
The instrument has an automatic switch-off function in order to save the
batteries (see section 11.6.1 S
, page 69). The automatic switchoff func-
tion switches off the meter if no key is pressed for an adjustable period.
The automatic switchoff function is not active
if the power pack is connected
if a USB-B cable is connected
if the
Automatic data storage
function is active, or with
automatic data trans-
mission
11.6.5 Display illumination
The meter automatically switches off the display illumination if no key is
pressed for 20 seconds.
The illumination is switched on with the next keystroke again.
You can also generally switch on the display illumination (see section 11.6.1
S
11.7 Reset
You can reset (initialize) all sensor settings and sensor-independent settings
separately from each other.
11.7.1 Resetting the measurement settings
pH
The following settings for pH measurements are reset to the default settings
with the
Reset
function:
The calibration data are reset to the default settings together with
the measuring parameters. Recalibrate after performing a reset.
Setting
Default settings
Buffer
TEC
Calibration interval
7 d
Unit for slope
mV/pH
Measured parameter
pH
Unit for slope
0.001
mV resolution
0.1
Asymmetry
0 mV