Technical data
MD 8000 L
12
ba77218e01
11/2018
Power pack
Helmsman Industrial Co Ltd SEI0901100P
Input: 100 ... 240 V ~ / 50 ... 60 Hz / 0.5 A
Output: 9 Vdc, 1100 mA
Connection max. overvoltage category II
ShenZhen RiHuiDa Power Supply Co Ltd
RHD10W090110
Input: 100 ... 240 V ~ / 50 ... 60 Hz / 0.4 A
Output: 9 Vdc, 1100 mA
Primary plugs
Primary plugs contained in the scope of
delivery:
Euro, US, UK and Australian.
USB interface (device)
Type
USB 1.1
USB-B (Device), PC
Baud rate
Adjustable:
1200, 2400, 4800, 9600, 19200 Baud
Data bits
8
Stop bits
2
Parity
None
Handshake
RTS/CTS
Cable length
max. 3 m (9.843 feet)
Guidelines
and norms used
EMC
EC directive 2004/108/EC
EN 61326-1
EN 61000-3-2
EN 61000-3-3
FCC Class A
Meter safety
EC directive 2006/95/EC
EN 61010-1
IP protection class
EN 60529