Settings
MD 8000 L
64
ba77218e01
11/2018
11.1.3 Calibration interval
The calibration evaluation is displayed as a sensor symbol.
Setting the calibra-
tion interval
The calibration interval is set to 7 days in the factory.
You can change the interval (1 ... 999 days):
11.2 ORP measurement settings
11.2.1 Settings for ORP measurements
The settings are made in the menu for measuring settings of the ORP measure-
ment. To open the settings, display the required measured parameter in the
measured value display and press the
<OK>
key. After completing the set-
tings, switch to the measured value display with
<MODE>
.
Default settings are printed in
bold
.
11.3 D.O. measurement settings
11.3.1 Settings for D.O. sensors
(menu for measurement and calibration settings)
Settings
The settings are available in the menu for measurement and calibration set-
tings. To open the settings, display the required measured parameter in the
measured value display and press the
<OK>
key. After completing the set-
To ensure the high measuring accuracy of the measuring system,
calibrate after the calibration interval has expired.
1.
Open the menu for measurement settings with
<OK>
.
2.
In the
Calibration
/
Calibration interval
menu, set the calibration interval
with
<
><
>
.
3.
Confirm the setting with
<OK>
.
4.
Quit the menu with
<MODE>
.
Menu item
Possible setting
Explanation
mV resolution
0.1
1
Resolution of the mV display
Reset
-
Resets all sensor settings to the
delivery condition (see section
11.7.1 R