Conductivity
MD 8000 L
54
ba77218e01
11/2018
Calibration evalua-
tion
After calibration, the meter automatically evaluates the current status of the cal-
ibration. The evaluation appears on the display and in the calibration record.
Calibration record
(
USB output
)
Display
Calibration record
Cell constant [cm
-1
]
+++
Within the range
0.450 ... 0.500 cm
-1
Error
Error
Outside the range
0.450 ... 0.500 cm
-1
Error elimination (see section 15
W
CALIBRATION Cond
01.11.2018 07:43:33
IDP 761-C
Ser. no. 09250033
Cell constant 0.476 1/cm 25.0 °C
Sensor
+++