9
vwr.com I Instruction manual VWR® A Water System
1.5 Operation Principles
VWR A water system produces ultrapure water and pure water directly from tap water. Most
particles, ions and organic compounds are removed through the pre-filtration, RO membrane.
When in need of ultrapure (UP) water, RO water goes through a UV lamp chamber with a dual
wavelength UV lamp to destroy trace organic pollutants in water, then through a polishing cartridge
to remove the last trace of ions, then a 0.2
μm final filter to the outlet.
VWR A Water Flow Diagram
No.
Description
No.
Description
C1
RO Booster Pump
C2
Distribution Pump
I1
Feed Conductivity Sensor
I2
RO Conductivity Sensor
I4
TOC Sensor (optional)
I5
UP Resistivity Sensor
N2
185 nm UV Lamp
O2
P Pack
O3
U Pack
O6
RO Packs
P
Pressure Sensor
S1
Inlet Filter
S2
Inline Filter
V
Flow Meter
S1
S2
I1
I2
I4
I5
O2
O6
O3
C1
C2
V
HP Dispenser
UP Dispenser
N2
Tap water