| Instruction manual
26
IT406 FL4 / IT406 FL4D
13.3 Installing fluorescence filter
Only IT406 FL4 / IT406 FL4D
This procedure can be needed when installing new fluorescence
cubes or when replacing an existing cube with a different one.
1
Disconnect the power supply plug from the microscope.
2
Unscrew the filter selector lever.
3
Open the side cover of the illuminator, by unscrewing the side
screws (1). (Fig. 42)
4
Remove the filter slider from the microscope and put it on a table.
5
Using the screws provided with the fluorescence cube, fix the new
cube on the slider. (Fig. 43)
1
2
FIGURE 42
FIGURE 43
Filter name
Excitation filter
Dichroic mirror
Barrier filter
Applications
VisiCube IT_B
325-375 nm
415 nm
435LP nm
–
DAPI
–
Hoechst
VisiCube IT_G
390-420 nm
440 nm
450LP nm
–
Coumarin
–
Pacific Blue
VisiCube IT_V
455-495 nm
500 nm
510LP nm
–
FITC
–
GFP
VisiCube IT_UV
510-550 nm
570 nm
575LP nm
–
Rhodamine
–
TRITC
–
Propidium Iodide
VisiCube IT_R1
582-603 nm
610 nm
615-645 nm
–
Red#2 FISH
VisiCube IT_R2
590-650 nm
660 nm
665LP nm
–
Cy5
VisiCube IT_R3
595-645 nm
655 nm
665-715 nm
–
Cy5
–
AlexaFluor 647
–
Draq5
VisiCube IT_FR
623-678 nm
685 nm
690-750 nm
–
Cy5.5
VisiCube IT_A
720-760 nm
770 nm
780LP nm
–
Li_Cor IR dye 800