Air Jacketed CO2 Incubator
7-3
VWR International
Section 7
Specifications
Weight
Net . . . . . . . . . . . . . . . . . . . . . . . . .260 lbs (118kg)
Net shipping . . . . . . . . . . . . . . . . .315 lbs (143kg)
Certifications
The Declaration of Conformity is available on request.
Safety Specifications
Indoor Use Only
Altitude . . . . . . . . . . . . . . . . . . . . . . .2,000 meters
Temperature . . . . . . . . . . . . . . . . . . . .5°C to 40°C
Humidity . . . . . . . . . . .80% RH at or below 31°C,
. . . . . . . .decreasing linearly to 50% RH at 40°C
Mains Supply Fluctuations
Mains supply voltage fluctuations not to exceed ±10%
of nominal voltage
Installation Category II
1
Pollution Degree 2
2
Class of Equipment
I
__________________________________________________________
1 Installation category (overvoltage category) defines the level of transient overvoltage
which the instrument is designed to withstand safely. It depends on the nature of the
electricity supply and its overvoltage protection means. For example, in CAT II which is
the category used for instruments in installations supplied from a supply comparable to
public mains such as hospital and research laboratories and most industrial laboratories,
the expected transient overvoltage is 2500V for a 230V supply and 1500V for a 120V
supply.
2 Pollution Degree describes the amount of conductive pollution present in the
operating environment. Pollution degree 2 assumes that normally only non-conductive
pollution such as dust occurs with the exception of occasional conductivity caused by
condensation.