background image

MULTI GAS LEAK DETECTOR

 

 

 

 

ME-LOKATOR VL30 – REFRIGERANT GASES

 

                                                                                                          

PAGE  |  

14

 

LOKATOR VL 30   EN 

ERROR & FAULT MESSAGES 

 

THE DETECTOR DOES NOT 

START 

When the push button is pressed for at least 2 seconds, the green start LED will not light and the detector will not 

start. 

 

Check that the power cord is firmly connected to the back of the cabinet and that the switch is in the ON (-) position. 

 

Check that the red power indicator on the front of the cabinet is lit. [Chap. 3.2.1] 

> If this is not the case, replace the fuse in the power supply unit on the back of the cabinet. [Chap. 3.2.2] 

 

If the power indicator light is on, check that you have pressed the power button for the right long time (approx. 2 seconds). 

> The green start indicator light should light on. 

 

If the detector still does not start, perform a RESET. 

> Use a rod (1mm diameter) and press the push button through the RESET hole. [Chap. 3.2.1] 

 

Try again to start the detector by pressing the power button. 

> The green start indicator light should light on. 

 

If the detector still does not start, contact the After Sales Service. 

 

SHORT CIRCUIT 

The display shows SHORT CIRCUIT. [Chap. 5.5] and the red "AUTO CONTROL" indicator flashes. 

 

Check that the cable is in good condition, and does not show signs of cuts, tears or crushing. It may exist in a cable in bad 

condition SHORT CIRCUITS. 

> Replace the cable if necessary then select "RESTART". 

Be careful using a faulty cable can seriously damage the detector. 

 

If the  SHORT CIRCUIT message persists, the  problem may come from the Sensor  Cell or Suction Block that is being 

damaged. 

> You may need to replace the Cell or Aspiration Block. 

 

If the SHORT CIRCUIT message still displayed, contact the After Sales Service. 

 

 

PROBE ERROR 

The screen displays PROBE ERROR. [Chap. 5.5] and the red "AUTO CONTROL" indicator flashes. 

 

Stop the detector by pressing the start button (2s approx.) 

 

Check the right connection of the sensor unit to the suction unit [chap. 4.1.1] 

 

Check that the cable is in good condition, and does not show signs of cuts, tears or crushing. A cable in poor condition may 

have WIRES CUT INSIDE. 

> Replace the cable if necessary. 

 

Restart the detector by pressing the start button (approx. 2s) 

 

If the ERROR PROBE message persists, the most common cause is that the Sensor Cell is worn or damaged. 

> It is necessary to replace the Cell of detection. 

 

If the ERROR PROBE message persists, contact the After Sales Service. 

 

 

LOW SENSITIVITY 

 

OR 

 

NO DETECTION 

After checking on a Calibrated Leak of known value, the detector detects little or no. 

Check that the red "AUTO CHECK" indicator is off. 

Check that the "CONTAMINATION" indicator is off. 

> If the light is on, the ambient working atmosphere is too polluted. Ventilate the work area and wait for the indicator to 

go out. 

> If the light does not go out, either the work area is still polluted or the Sensor Cell is over used and it is necessary to 

replace it. 

Check the detector settings: GAS Type and DETECTION Threshold. 

Check the CALIBRATION of the detector [chap. 5.4.1] 

> If necessary, make a new calibration with a Calibrated Leak of known value. Otherwise, it is possible to restore the 

factory settings of the detector. [Chap. 6.4.3] 

> If calibration is not possible, either check your Calibrated Leak or the Sensor Cell is used and it is necessary to replace 

it. 

Unscrew the suction block from the cable. 

Unscrew the sniffer pipe from the suction block. 

Check that the sniffing pipe is not clogged, bent or crushed. 

> If this is the case try to unclog the pipe, otherwise it is necessary to replace the sniffing pipe. 

Check that the cell is equipped with a fiber filter and that it is not obstructed or room. 

> If this is the case replace the fiber filter. 

Check that the teflon end of the sniffer pipe is not clogged. 

> If this is the case try to unclog the Teflon tip with a tipor replace it. 

10 

Check that you hear the sound of the suction turbine in the suction unit. 

He must be gentle and steady. 

> In the event that the turbine does not rotate, emits whistles or is irregular in operation, it is necessary to replace the 

SUCTION BLOCK. 

 

11 

The most common cause of low sensitivity or inability to detect is a used detection cell. 

> It is necessary to replace the detection cell. 

 

If the problem persists, contact the After Sales Service. 

 

 

Summary of Contents for Lokring LOKATOR VL30

Page 1: ...LOKATOR VL30 MULTI GAS LEAK DETECTOR INSTRUCTION MANUAL LOKATOR VL30 REF REV 1 0...

Page 2: ...DETECTION LIGHT RAMP and SOUND SIGNAL 8 5 2 2 DETECTION SCREEN 9 5 3 LEAK REASEARCH DETECTION 10 5 3 1 LEAK RESEARCH PROCEDURE 10 5 3 2 ZEROTTING STABILITY 10 5 3 3 SATURATION CONTAMINATION CONTROLE...

Page 3: ...ing phenomena on the gas sensor that may alter the sensitivity and or selectivity of the sensor Reduced sensitivity unstable calibration or false detection may be a possible consequence Commissioning...

Page 4: ...mer Adjustable sound volume Indication of environmental contamination SELF CHECKING Permanent self checking Self checking of the Detection Cell Short circuit protection in the double plug cable COMMUN...

Page 5: ...mains power socket 110v to 250v 50 60Hz 3 2 3 LEFT PANEL OF ELECTRONIC UNIT 1 Logic output DB9 socket 1 Ready to Detect 2 Busy 3 Error 4 Detection 5 Detec Threshold 6 Contamination 7 N A 8 GND 0v 9 12...

Page 6: ...ows The detection of all the refrigerant gases currently used HFC HFO and Mixture Detection of the R600 R290 MINIMUM SENSITIVITY 1234ze 0 3 g yr MINIMUM SENSITIVITY R134a 0 3 g yr MINIMUM SENSITIVITY...

Page 7: ...50v 50 60 Hz Earth Mains Plug and fuse of 500 mA are located on the back of the cabinet consumption 25 W approx 4 1 4 OPERATION WITH INCORPORATED BATTERIES PACK As an option the LOKATOR VL30 Detector...

Page 8: ...he front panel of the cabinet lights up to indicate Mains power Wait 15 seconds and press the push button on the front panel for at least 2 seconds The green LED on the front panel lights up to indica...

Page 9: ...efully do not hit it do not knock it down A circular seal protects the block head against shocks At rest hang the detection block on the support provided with the detector A fiber filter after the sem...

Page 10: ...N CONTAMINATION of the cell x10 5 2 2 DETECTION SCREEN 1 Profile gas in use ex Profile A R32 2 Profile detection threshold in use 3 CALIBRATION Menu 4 Battery level display BATTERY Option 5 Select and...

Page 11: ...e up the shape required by the user However avoid bending the tube into excessively sharp angles 5 3 2 ZEROTTING STABILITY With a new Sensor the manipulation of the Detection Probe can cause the zero...

Page 12: ...AKEN BEFORE CALIBRATION PROCEDURE If the calibrated leak is adjustable pressure check the pressure corresponding to the calibration value During the calibration procedure move the probe smoothly Posit...

Page 13: ...the detector 6 OPTIONS MENU The LOKATOR VL30 has 4 options screen OPTION 1 OPTION 2 OPTION 3 and OPTION 4 It is accessed by selecting OPTIONS 6 1 OPTION 1 PROFILE A SELECTION OF THE TYPE OF GAS SET TO...

Page 14: ...ION Set the maximum degassing time of the LOKATOR VL30 after detection OFF the degassing time depends on the leak detected Degassing from 0s to 60s max 5s 30s the maximum degassing time is that set De...

Page 15: ...cable in poor condition may have WIRES CUT INSIDE Replace the cable if necessary 4 Restart the detector by pressing the start button approx 2s 5 If the ERROR PROBE message persists the most common cau...

Page 16: ...ction pipe The fiber filter mounted at the entrance to the Detection of Cell is replaceable by unscrewing the sniffing pipe This filter must be changed regularly and as soon as necessary 8 3 PERIODIC...

Page 17: ...MULTI GAS LEAK DETECTOR ME LOKATOR VL30 REFRIGERANT GASES PAGE 16 LOKATOR VL 30 EN 10 CE DECLARATION...

Page 18: ...7 LOKATOR VL 30 EN 11 TECHNICAL DATA Warm up time 180 sec Reaction time 1 sec Operating temperature 10 C 40 C Interface DB9 Logical signal output 0 V 5 V Power supply 110 V 250 V 50 60 Hz Power consum...

Reviews: