5.4 

CALIBRATION FUNCTION 

 
Access  to  the  CALIBRATION  function  by  selecting  on  the  main 
DETECTION screen "CAL.[chap. 5.2.2 

N°3

] 

 

 
During its use, the sensitivity of the detection cell may vary depending on 
various factors such as the frequency of detection, ambiant contamination 
and other soiling factors. 

 

 

• Before the start of a CALIBRATION procedure, the probe should be at least in operation for 10 minutes minimum 

and in a atmosphere not or very little contaminated to ensure optimum calibration. 

 

• The CALIBRATION procedure must be carried out in a room with a low airflow. 

 

• A CALIBRATION on a leakage value less than 3 g/yr must be ONLY performed in an environment without NO 

CONTAMINATION. 

 

 

5.4.1 

CHECK OF CALIBRATION 

 

• Use a calibrated leak according the gas of active profile A or B [chap. 6] and set its value. 

 

-

 

CALIBRATED Leak R134a    

0,5 to 8 g/yr  

-

 

CALIBRATED Leak 1234ze    

0,5 to 8 g/yr 

-

 

CALIBRATED Leak R32  

 

0,5 to 8 g/yr 

-

 

CALIBRATED Leak R290  

 

0,5 to 8 g/yr 

-

 

CALIBRATED Leak R410  

 

0,5 to 8 g/yr 

-

 

CALIBRATED Leak R600  

 

0,5 to 8 g/yr 

-

 

CALIBRATED Leak CUSTOM   

0,5 to 8 g/yr 

 

• Select "TEST" to start a calibration check procedure. 
• At the end of the check, the display indicates whether the TEST CALIBRATION is "OK" or "FALSE". 
 
• Following a TEST CALIBRATION, if the TEST is "FALSE", initiate a CALIBRATION procedure. 

  

5.4.2 

PRECAUTIONS TO BE TAKEN BEFORE CALIBRATION PROCEDURE 

 

• If the calibrated leak is adjustable pressure, check the pressure corresponding to the calibration value. 
• During the calibration procedure, move the probe smoothly. 
• Position the nozzle of the probe pipe directly into the orifice of the calibrated leak.  

 

5.4.3 

CALIBRATION PROCEDURE 

 

Use a calibrated leak according the gas of active profile A or B [chap. 6], set its value and then press the CALIBRATION button: 

 

-

 

CALIBRATED Leak R134a    

0,5 to 8 g/yr  

-

 

CALIBRATED Leak 1234ze    

0,5 to 8 g/yr 

-

 

CALIBRATED Leak R32  

 

0,5 to 8 g/yr 

-

 

CALIBRATED Leak R290  

 

0,5 to 8 g/yr 

-

 

CALIBRATED Leak R410  

 

0,5 to 8 g/yr 

-

 

CALIBRATED Leak R600  

 

0,5 to 8 g/yr 

-

 

CALIBRATED Leak CUSTOM   

0,5 to 8 g/yr 

 

CUSTOM gas: The LOKATOR VL30 has 2 CUSTOM gases in order to calibrate the detector on a gas not included in the library. 

-

 

CUST.1> Position for R134a type gases, ... 

-

 

CUST.2> Position for gases type R410, R32, R600, ... 

Refer to the after sales service to guide you on the best CUSTOM position to use. 

 
• Then follow the different steps of the CALIBRATION protocol displayed on the screen. 
• At any time, you can cancel the CALIBRATION procedure by selecting  « CANCEL ». 
• If the CALIBRATION protocol ran without errors, the new calibration value is saved automatically. 
• It is possible to reset the calibration value to the factory calibration value in the OPTION menu of the LOKATOR VL30. 

[chap. 6.4.3] 

 

If the message CALIBRATION FAIL is displayed, it is possible that the flow of the calibrated leak is too low or that the 
sensing cell is desensitized; making the calibration impossible. 
Check the calibrated leak or the sensor cell.

 

 

 

Summary of Contents for Lokring LOKATOR VL30

Page 1: ...LOKATOR VL30 MULTI GAS LEAK DETECTOR INSTRUCTION MANUAL LOKATOR VL30 REF REV 1 0...

Page 2: ...DETECTION LIGHT RAMP and SOUND SIGNAL 8 5 2 2 DETECTION SCREEN 9 5 3 LEAK REASEARCH DETECTION 10 5 3 1 LEAK RESEARCH PROCEDURE 10 5 3 2 ZEROTTING STABILITY 10 5 3 3 SATURATION CONTAMINATION CONTROLE...

Page 3: ...ing phenomena on the gas sensor that may alter the sensitivity and or selectivity of the sensor Reduced sensitivity unstable calibration or false detection may be a possible consequence Commissioning...

Page 4: ...mer Adjustable sound volume Indication of environmental contamination SELF CHECKING Permanent self checking Self checking of the Detection Cell Short circuit protection in the double plug cable COMMUN...

Page 5: ...mains power socket 110v to 250v 50 60Hz 3 2 3 LEFT PANEL OF ELECTRONIC UNIT 1 Logic output DB9 socket 1 Ready to Detect 2 Busy 3 Error 4 Detection 5 Detec Threshold 6 Contamination 7 N A 8 GND 0v 9 12...

Page 6: ...ows The detection of all the refrigerant gases currently used HFC HFO and Mixture Detection of the R600 R290 MINIMUM SENSITIVITY 1234ze 0 3 g yr MINIMUM SENSITIVITY R134a 0 3 g yr MINIMUM SENSITIVITY...

Page 7: ...50v 50 60 Hz Earth Mains Plug and fuse of 500 mA are located on the back of the cabinet consumption 25 W approx 4 1 4 OPERATION WITH INCORPORATED BATTERIES PACK As an option the LOKATOR VL30 Detector...

Page 8: ...he front panel of the cabinet lights up to indicate Mains power Wait 15 seconds and press the push button on the front panel for at least 2 seconds The green LED on the front panel lights up to indica...

Page 9: ...efully do not hit it do not knock it down A circular seal protects the block head against shocks At rest hang the detection block on the support provided with the detector A fiber filter after the sem...

Page 10: ...N CONTAMINATION of the cell x10 5 2 2 DETECTION SCREEN 1 Profile gas in use ex Profile A R32 2 Profile detection threshold in use 3 CALIBRATION Menu 4 Battery level display BATTERY Option 5 Select and...

Page 11: ...e up the shape required by the user However avoid bending the tube into excessively sharp angles 5 3 2 ZEROTTING STABILITY With a new Sensor the manipulation of the Detection Probe can cause the zero...

Page 12: ...AKEN BEFORE CALIBRATION PROCEDURE If the calibrated leak is adjustable pressure check the pressure corresponding to the calibration value During the calibration procedure move the probe smoothly Posit...

Page 13: ...the detector 6 OPTIONS MENU The LOKATOR VL30 has 4 options screen OPTION 1 OPTION 2 OPTION 3 and OPTION 4 It is accessed by selecting OPTIONS 6 1 OPTION 1 PROFILE A SELECTION OF THE TYPE OF GAS SET TO...

Page 14: ...ION Set the maximum degassing time of the LOKATOR VL30 after detection OFF the degassing time depends on the leak detected Degassing from 0s to 60s max 5s 30s the maximum degassing time is that set De...

Page 15: ...cable in poor condition may have WIRES CUT INSIDE Replace the cable if necessary 4 Restart the detector by pressing the start button approx 2s 5 If the ERROR PROBE message persists the most common cau...

Page 16: ...ction pipe The fiber filter mounted at the entrance to the Detection of Cell is replaceable by unscrewing the sniffing pipe This filter must be changed regularly and as soon as necessary 8 3 PERIODIC...

Page 17: ...MULTI GAS LEAK DETECTOR ME LOKATOR VL30 REFRIGERANT GASES PAGE 16 LOKATOR VL 30 EN 10 CE DECLARATION...

Page 18: ...7 LOKATOR VL 30 EN 11 TECHNICAL DATA Warm up time 180 sec Reaction time 1 sec Operating temperature 10 C 40 C Interface DB9 Logical signal output 0 V 5 V Power supply 110 V 250 V 50 60 Hz Power consum...

Reviews: