background image

MULTI GAS LEAK DETECTOR

 

 

 

 

ME-LOKATOR VL30 – REFRIGERANT GASES

 

                                                                                                          

PAGE  |  

10

 

LOKATOR VL 30   EN 

5.3 

LEAK REASEARCH & DETECTION

 

 

5.3.1 

LEAK RESEARCH PROCEDURE

 

 

 The response time for detection is approx. 1 second. 

The response time depends on the length of the sniffing tube used. 

 

 To obtain a quick response and an efficient detection, the sniffing tube must be brought as close as possible to 

the place being tested. (A few millimeters) 

 

 Avoid holding the sniffing tube on the leakage point excessively, as detector sensor fatigue will result. 

 

 Always work by successive contacts, pulling back the tube from the leak point when a detection. 
 
 
STEP 1 >
  

When the warm-up period over, the red light goes out and the “0” green light of the light ramp is ON 
indicating that the unit is ready for detection. 

 
STEP 2 >  

 Check leak potential points. 
 Always work by successive contacts, pulling back the tube from the leak point when a detection. 

 
STEP 3 >
  

 On pulling back the detection pipe from the leakage point, the YELLOW 

“DEGASSING” light on. The GREEN “0” light will light ON during the degassing time of the sensor. 

 

It is imperative to wait for the YELLOW "DEGASSING" light to go out before continuing the detection. 
 
 The sniffing tube is semi-rigid and can take up the shape required by the user. However, avoid bending the tube into 

excessively sharp angles. 

 
 

5.3.2 

ZEROTTING STABILITY

 

 

• With a new Sensor the manipulation of the Detection Probe can cause the zero to be unstable, especially in high 

sensitivity positions. This instability will normally disappear after a few uses. 

 
• Avoid manipulating the Detection Probe roughly and knocking it against the object being tested. 
 
• Atmospheric contamination (Gases, Refrigerant gases or Solvents) at an excessive level can also cause zero instability 

and lead to false readings. 

 
• Atmospheric contamination at an excessive level cannot be compensated for by the auto-zero and will produce a 

permanently illuminated display (especially when using the positions of high sensitivity). 

 
 

5.3.3 

 « SATURATION - CONTAMINATION » CONTROLE 

 

The YELLOW "x10" "CONTAMINATION" light at the right end of the light ramp is normally off. This light comes on for two 
reasons: 
 
• The saturation of the Sensor when detecting a very big leak. 

In that case it is necessary to wait for the light to be fully OFF before continuing the detection. 

 
• A very high atmospheric contamination can makes difficult or impossible an efficient tightness control and will cause 

a premature wear of the Sensor. In such a case, it is essential to reduce the atmospheric contamination level by 
ventilation of the premises, and to stop all sources of contamination before continuing the detection. 

 

 

Summary of Contents for Lokring LOKATOR VL30

Page 1: ...LOKATOR VL30 MULTI GAS LEAK DETECTOR INSTRUCTION MANUAL LOKATOR VL30 REF REV 1 0...

Page 2: ...DETECTION LIGHT RAMP and SOUND SIGNAL 8 5 2 2 DETECTION SCREEN 9 5 3 LEAK REASEARCH DETECTION 10 5 3 1 LEAK RESEARCH PROCEDURE 10 5 3 2 ZEROTTING STABILITY 10 5 3 3 SATURATION CONTAMINATION CONTROLE...

Page 3: ...ing phenomena on the gas sensor that may alter the sensitivity and or selectivity of the sensor Reduced sensitivity unstable calibration or false detection may be a possible consequence Commissioning...

Page 4: ...mer Adjustable sound volume Indication of environmental contamination SELF CHECKING Permanent self checking Self checking of the Detection Cell Short circuit protection in the double plug cable COMMUN...

Page 5: ...mains power socket 110v to 250v 50 60Hz 3 2 3 LEFT PANEL OF ELECTRONIC UNIT 1 Logic output DB9 socket 1 Ready to Detect 2 Busy 3 Error 4 Detection 5 Detec Threshold 6 Contamination 7 N A 8 GND 0v 9 12...

Page 6: ...ows The detection of all the refrigerant gases currently used HFC HFO and Mixture Detection of the R600 R290 MINIMUM SENSITIVITY 1234ze 0 3 g yr MINIMUM SENSITIVITY R134a 0 3 g yr MINIMUM SENSITIVITY...

Page 7: ...50v 50 60 Hz Earth Mains Plug and fuse of 500 mA are located on the back of the cabinet consumption 25 W approx 4 1 4 OPERATION WITH INCORPORATED BATTERIES PACK As an option the LOKATOR VL30 Detector...

Page 8: ...he front panel of the cabinet lights up to indicate Mains power Wait 15 seconds and press the push button on the front panel for at least 2 seconds The green LED on the front panel lights up to indica...

Page 9: ...efully do not hit it do not knock it down A circular seal protects the block head against shocks At rest hang the detection block on the support provided with the detector A fiber filter after the sem...

Page 10: ...N CONTAMINATION of the cell x10 5 2 2 DETECTION SCREEN 1 Profile gas in use ex Profile A R32 2 Profile detection threshold in use 3 CALIBRATION Menu 4 Battery level display BATTERY Option 5 Select and...

Page 11: ...e up the shape required by the user However avoid bending the tube into excessively sharp angles 5 3 2 ZEROTTING STABILITY With a new Sensor the manipulation of the Detection Probe can cause the zero...

Page 12: ...AKEN BEFORE CALIBRATION PROCEDURE If the calibrated leak is adjustable pressure check the pressure corresponding to the calibration value During the calibration procedure move the probe smoothly Posit...

Page 13: ...the detector 6 OPTIONS MENU The LOKATOR VL30 has 4 options screen OPTION 1 OPTION 2 OPTION 3 and OPTION 4 It is accessed by selecting OPTIONS 6 1 OPTION 1 PROFILE A SELECTION OF THE TYPE OF GAS SET TO...

Page 14: ...ION Set the maximum degassing time of the LOKATOR VL30 after detection OFF the degassing time depends on the leak detected Degassing from 0s to 60s max 5s 30s the maximum degassing time is that set De...

Page 15: ...cable in poor condition may have WIRES CUT INSIDE Replace the cable if necessary 4 Restart the detector by pressing the start button approx 2s 5 If the ERROR PROBE message persists the most common cau...

Page 16: ...ction pipe The fiber filter mounted at the entrance to the Detection of Cell is replaceable by unscrewing the sniffing pipe This filter must be changed regularly and as soon as necessary 8 3 PERIODIC...

Page 17: ...MULTI GAS LEAK DETECTOR ME LOKATOR VL30 REFRIGERANT GASES PAGE 16 LOKATOR VL 30 EN 10 CE DECLARATION...

Page 18: ...7 LOKATOR VL 30 EN 11 TECHNICAL DATA Warm up time 180 sec Reaction time 1 sec Operating temperature 10 C 40 C Interface DB9 Logical signal output 0 V 5 V Power supply 110 V 250 V 50 60 Hz Power consum...

Reviews: