MULTI GAS LEAK DETECTOR

 

 

 

 

ME-LOKATOR VL30 – REFRIGERANT GASES

 

                                                                                                          

PAGE  |  

13

 

LOKATOR VL 30   EN 

6.2.1 

SELECTION OF THE TYPE OF GAS SET TO PROFILE B 

 

Selects the type of gas to be detected assigned to PROFILE B. The LOKATOR VL30 will automatically select the right calibration 
according to the type of gas. 
 

6.2.2 

SELECTION OF THE THRESHOLD SET TO PROFILE B 

 

Set the value of the detection threshold assigned to the PROFILE B. 
To optimize the accuracy reading of the leakage value, it is preferable to calibrate the detector with a calibrated leak with a 
value close to the selected profile detection threshold. 

 

 

6.3 

OPTION 3

 

 

 

 

CURRENT UNIT [6.3.1]

 

 

LANGUAGE SETTING [6.3.2]

 

 

« BACK » return to the screen « OPTION 2 

» 

 

 

« NEXT » access to the screen « OPTION 4 

» 

 

 
 

6.3.1 

CURRENT UNIT 

 

The LOKATOR VL 30 version REFIGERANT GASES displays leakage values only in [G/YR] 
 

6.3.2 

LANGUAGE SETTING 

 

Used to set the language of the LOKATOR VL30 : [FR] French – [EN] English 

 

 

6.4 

OPTION 4

 

 

 

FAST DEGASSING OPTION [6.4.1]

 

 

AUTO-START OPTION [6.4.1]

 

 

RESET – FACTORY SETTING [6.4.3]

 

 

« BACK » return to the screen « OPTION 2 »  

 

 

6.4.1 

FAST DEGASSING OPTION 

 

Set the maximum degassing time of the LOKATOR VL30 after detection. 

 

• [OFF]         > the degassing time depends on the leak detected - Degassing from 0s to 60s max. 

 

• [5s – 30s] > the maximum degassing time is that set - Degassing from 5s to 30s max. 

 

By activating the FAST DEGASSING option, if large leaks are detected, the degassing time of the cell will not be 
sufficient. It may be that for a few seconds, the following detections will be made with a loss of sensitivity. 

 

6.4.2 

AUTO-START OPTION 

 

Allows the LOKATOR VL30 to start automatically when powered by the mains. This makes it possible to use the LOKATOR 
VL30 with a general power supply from a workstation or controlled by an automatic system. 

 

• [ON]  

> the LOKATOR VL30 starts automatically as soon as it is powered by the mains. To turn it off, switch off the main 
or press at least 2 seconds on the start push button. 
[chap. 3.2.1 

n°9

] 

 

• [OFF]   > the automatic start of the LOKATOR VL30 is deactivated. To start it, you have to press at least 2 seconds on the 

start push button and to shut down press again 2 seconds on push button. 
[chap. 3.2.1 

n°9

 

6.4.3 

RESET – FACTORY SETTING 

 

Resets the LOKATOR VL30. All settings and the CALIBRATION value will be reset to their default values. 
< BACK return to the screen « OPTION 4 »  

 
 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Summary of Contents for Lokring LOKATOR VL30

Page 1: ...LOKATOR VL30 MULTI GAS LEAK DETECTOR INSTRUCTION MANUAL LOKATOR VL30 REF REV 1 0...

Page 2: ...DETECTION LIGHT RAMP and SOUND SIGNAL 8 5 2 2 DETECTION SCREEN 9 5 3 LEAK REASEARCH DETECTION 10 5 3 1 LEAK RESEARCH PROCEDURE 10 5 3 2 ZEROTTING STABILITY 10 5 3 3 SATURATION CONTAMINATION CONTROLE...

Page 3: ...ing phenomena on the gas sensor that may alter the sensitivity and or selectivity of the sensor Reduced sensitivity unstable calibration or false detection may be a possible consequence Commissioning...

Page 4: ...mer Adjustable sound volume Indication of environmental contamination SELF CHECKING Permanent self checking Self checking of the Detection Cell Short circuit protection in the double plug cable COMMUN...

Page 5: ...mains power socket 110v to 250v 50 60Hz 3 2 3 LEFT PANEL OF ELECTRONIC UNIT 1 Logic output DB9 socket 1 Ready to Detect 2 Busy 3 Error 4 Detection 5 Detec Threshold 6 Contamination 7 N A 8 GND 0v 9 12...

Page 6: ...ows The detection of all the refrigerant gases currently used HFC HFO and Mixture Detection of the R600 R290 MINIMUM SENSITIVITY 1234ze 0 3 g yr MINIMUM SENSITIVITY R134a 0 3 g yr MINIMUM SENSITIVITY...

Page 7: ...50v 50 60 Hz Earth Mains Plug and fuse of 500 mA are located on the back of the cabinet consumption 25 W approx 4 1 4 OPERATION WITH INCORPORATED BATTERIES PACK As an option the LOKATOR VL30 Detector...

Page 8: ...he front panel of the cabinet lights up to indicate Mains power Wait 15 seconds and press the push button on the front panel for at least 2 seconds The green LED on the front panel lights up to indica...

Page 9: ...efully do not hit it do not knock it down A circular seal protects the block head against shocks At rest hang the detection block on the support provided with the detector A fiber filter after the sem...

Page 10: ...N CONTAMINATION of the cell x10 5 2 2 DETECTION SCREEN 1 Profile gas in use ex Profile A R32 2 Profile detection threshold in use 3 CALIBRATION Menu 4 Battery level display BATTERY Option 5 Select and...

Page 11: ...e up the shape required by the user However avoid bending the tube into excessively sharp angles 5 3 2 ZEROTTING STABILITY With a new Sensor the manipulation of the Detection Probe can cause the zero...

Page 12: ...AKEN BEFORE CALIBRATION PROCEDURE If the calibrated leak is adjustable pressure check the pressure corresponding to the calibration value During the calibration procedure move the probe smoothly Posit...

Page 13: ...the detector 6 OPTIONS MENU The LOKATOR VL30 has 4 options screen OPTION 1 OPTION 2 OPTION 3 and OPTION 4 It is accessed by selecting OPTIONS 6 1 OPTION 1 PROFILE A SELECTION OF THE TYPE OF GAS SET TO...

Page 14: ...ION Set the maximum degassing time of the LOKATOR VL30 after detection OFF the degassing time depends on the leak detected Degassing from 0s to 60s max 5s 30s the maximum degassing time is that set De...

Page 15: ...cable in poor condition may have WIRES CUT INSIDE Replace the cable if necessary 4 Restart the detector by pressing the start button approx 2s 5 If the ERROR PROBE message persists the most common cau...

Page 16: ...ction pipe The fiber filter mounted at the entrance to the Detection of Cell is replaceable by unscrewing the sniffing pipe This filter must be changed regularly and as soon as necessary 8 3 PERIODIC...

Page 17: ...MULTI GAS LEAK DETECTOR ME LOKATOR VL30 REFRIGERANT GASES PAGE 16 LOKATOR VL 30 EN 10 CE DECLARATION...

Page 18: ...7 LOKATOR VL 30 EN 11 TECHNICAL DATA Warm up time 180 sec Reaction time 1 sec Operating temperature 10 C 40 C Interface DB9 Logical signal output 0 V 5 V Power supply 110 V 250 V 50 60 Hz Power consum...

Reviews: