SOURce:SEQuence:LENGth
Purpose
Sets the number of segments in a sequence list.
Type
Setting
Command Syntax
[SOURce:]SEQuence:LENGth <numeric_value>
Command Parameters
<numeric_value> = A numeric value from 1 to 4096
*RST Value
1
Query Syntax
[SOURce:]SEQuence:LENGth?
Query Parameters
N/A
Query Response
Numeric value from 1 to 4096
Description
The Source Sequence Length command sets the number of segments in a sequence list. A
maximum of 4096 segments may be defined or a minimum of 1 may be used.
Examples
Command / Query
Response
(Description)
SEQ:LENG 20
SEQ:LENG?
20
Related Commands
SOURce:SEQuence:LIST