www.vtiinstruments.com
VM3608A/3616A Command Dictionary
101
SOURce:ROSCillator:SLOPe
Purpose
Selects the active slope for advancing data when in FIFO or ARBitrary waveform
generation mode.
Type
Setting
Command Syntax
SOURce:ROSCillator:SLOPe <slope>
Command Parameters
<slope> = POSitive | NEGative.
*RST Value
<slope> = POS
Query Syntax
SOURce:ROSCillator:SLOPe?
Query Parameters
N/A
Query Response
POS | NEG
Description
The Source Roscillator Slope command selects the active edge for advancing data when
in FIFO, or ARBitrary waveform generation mode. This Command Only applies to the
front panel clock input and to the TTL trigger bus inputs. If the positive edge is selected,
a rising edge on the clock signal will cause the data to advance. If the negative edge is
selected, a falling edge will advance the data. If the timer or internal clock source is
selected, this command will have no effect.
Examples
Command / Query
Response
(Description)
SOUR:ROSC:SLOP NEG
(Selects the falling edge)
SOUR:ROSC:SLOP?
NEG
(Queries then reports that the current
setting is a negative slope.)
Related Commands
Summary of Contents for VM3608A
Page 2: ...VTI Instruments Corp 2 ...
Page 6: ...VTI Instruments Corp 6 VM3608A 3616A Preface ...
Page 12: ...VTI Instruments Corp 12 VM3608A 3616A Preface ...
Page 18: ...VTI Instruments Corp 18 VM3608A 3616A Introduction ...
Page 134: ...VTI Instruments Corp 134 VM3608A 3616A Command Dictionary ...