VM3608A/3616A Programming
In this example, the VM3608A/3616A loads scan list of the specified channel with voltage values.
These values can be loaded to the DAC using the interrupt routine-Note, for the above to function,
the trigger source should be either EXTernal or TTLTrigger 0 - 7.
TRIG:SOUR EXT
Sets the trigger source as EXT.
SCAN 1, (@1:4)
Sets the scan mode ON for Channels 1 through 4.
SCAN:L1M 1,256
Sets the position in the scan list arrays for Channel 1 as 256.
SCAN:TAB 1,2,3,4,5
Loads the Channel 1’s scan list with voltage values 2, 3, 4 and
5.
In this example, the memory list of the VM3608A/3616A is setup and then the output levels of the
channels are programmed using the memory lists.
MEM:SETUP 1, 2,3,4,5,6,7,8,10
Setting up memory list of Channels 1 through 8 at index 1
assuming VM3608A.
MEM:SETUP 2, 10,11,12,14,15,12,11,1
Setting up memory list of Channels 1 through 8 at index 2
assuming VM3608A.
SOUR: VOLT: SETUP 1
Setting the DAC output levels to the voltages programmed in
memory lists at index 1.