www.vtiinstruments.com
EX1401 Instrument Drivers
75
Trigger Offset
(always zero if
TTI is used)
Trigger Delay
> 0
Trigger
First data point
Sample
Interval
InitialX
> 0
Waveform
Record
Trigger Holdoff
Trigger
Pre-Trigger
Samples
Re-arm Time
F
IGURE
5-2:
P
OSITIVE
T
RIGGER
D
ELAY
Trigger
Offset
Trigger Delay <
0
Trigger
First data point
Sample
Interval
InitialX <
0
Waveform Record
Pre-Trigger
Samples
Re-arm Time
Trigger Holdoff
F
IGURE
5-3:
N
EGATIVE
T
RIGGER
D
ELAY
The
TriggerCount
property defines how many times the trigger events are accepted and data blocks
are acquired. After “
TriggerCount
” triggers are processed, the measurement waits for the next
arming condition or finishes. The
Holdoff
time specifies the minimum amount of time the
measurement has to wait before it can be triggered again once a trigger is detected. Any trigger
Summary of Contents for EX1401
Page 2: ......
Page 24: ......
Page 57: ...www vtiinstruments com EX1401 Web Page Operation 57 ...
Page 94: ......
Page 98: ......
Page 102: ......
Page 111: ...www vtiinstruments com EX10xxA Index 111 FIGURE A 1 IEEE1588 NETWORK TOPOLOGY ...
Page 112: ......