EX1401 Introduction
23
Choose an appropriate sampling rate
For best instrument noise performance, the sampling rate should be set as low as the data collection
requirements allow. For more details, see
Select proper location for installation
The EX1401 unit should be located away from sources of high or low temperature, strong air
currents, and high magnetic fields.
Use correct wiring
Best results will be achieved with the shortest and thickest thermocouple wire that the physical
requirements of the application can support. In addition, shielded thermocouples can be employed
to raise the system’s rejection of electrical interference. For more details, see
in Section 2.