VTI Instruments Corp.
94
EX1200-4260 Plug-in Module
EX1200-4260
S
PECIFICATIONS
G
ENERAL
S
PECIFICATIONS
C
HANNEL
C
OUNT
Two 60 x 2 2-wire channels
R
ELAY
T
YPE
Electromechanically, fail-safe
M
AXIMUM
S
WITCHING
V
OLTAGE
220 V dc/ 250 V ac rms
M
AXIMUM
S
WITCHING
C
URRENT
2 A
M
AXIMUM
S
WITCHING
P
OWER
60 W, 62.5 VA (see the
366H
Figure 4- for more information)
M
INIMUM
C
ONTACT
R
ATING
*
100 µV
*This value is in reference to a resistive load. Minimum capacity changes depending on switching frequency and environmental conditions
R
ATED
S
WITCH
O
PERATIONS
Mechanical
Electrical
1 x 10
8
1 x 10
5
at full load
S
WITCHING
T
IME
(T
YPICAL
)
< 10 ms
P
ATH
R
ESISTANCE
< 0.5 Ω
I
NSULATION
R
ESISTANCE
>10
9
Ω
M
AXIMUM
T
HERMAL
O
FFSET
PER
C
HANNEL
(HI-LO)
< 10 μV
B
ANDWIDTH
(-3
dB)
> 40 MHz
C
ROSSTALK
(T
YPICAL
)
1 MHz
10 MHz
< -45 dB
< -30 dB
I
SOLATION
(T
YPICAL
)
1 MHz
10 MHz
< -60 dB
< -50 dB
For mating connector, crimp pins, and other accessories, please refer to
Appendix B
.
R
ELAY
B
REAKING
C
APACITY
F
IGURE
4-37:
R
ELAY
B
REAKING
C
APACITY
Summary of Contents for EX1200 Series