VTI Instruments Corp.
18
EX1200 Series Terminal Blocks
2)
Save and provide a name for the DMM configuration for the scanning purposes (here,
TEMP
is used).
3)
Setup the scanlist
a. Add a scan step with a sweep phase that scans CH1_1 using the previously saved
configuration (e.g. CH1_1=
TEMP
)
b. The setup phase can be ignored for this example.
4)
Initiate & Read
a. Initiate the scan
b. Read back the data
Use Thermistor as a Reference Junction
1)
Set the DMM to measure temperature using the thermistor.
a. Set the DMM function to
Temperature
.
b. Set the transducer type to
Thermistor
.
c. Set the thermistor resistance to
10000
(the thermistor is 10 kΩ).
d. Set the
isReferenceJunction
parameter to
True
.
2)
Save and provide a name for the DMM configuration for the scanning purposes.
3)
Set the DMM to measure temperature using a Type K thermocouple (any thermocouple type
can be selected here, but Type K below).
a. Set the DMM function to
Temperature
.
b. Set the transducer type to
Thermocouple
.
c. Set the thermocouple type to
Type K
.
d. Set the reference type to
External
.
e. Set the
isReferenceJunction
parameter to
False
.
4)
Save and provide a name for the DMM configuration for the scanning purposes.
5)
Setup the scanlist
a. Add a scan step with a sweep phase that scans CH1_1 using the first saved configuration.
(e.g. CH1_1=
RefJunction
).
b. Add another scan step with a sweep phase that scans another channel (e.g.
CH1_2=
TEMP
).
c. The setup phase can be ignored for this example.
6)
Initiate & Read
a. Initiate the scan
b. Read back the data