www.vtiinstruments.com
EX1200-3072 Plug-in Module
57
EX1200-3072
S
PECIFICATIONS
G
ENERAL
S
PECIFICATIONS
C
HANNEL
C
OUNT
(1 x 72) 2-wire, dual (1 x 36) 2-wire, or (1 x 36) 4-wire
R
ELAY
T
YPE
Electromechanical, fail-safe
M
AXIMUM
S
WITCHING
V
OLTAGE
300 V ac rms, 300 V dc
M
AXIMUM
S
WITCHING
C
URRENT
2 A
M
AXIMUM
S
WITCHING
P
OWER
60 W dc, 125 VA
*Maximum switched power is at 30 V/ 2 A dc. Max switched power is derated non-linearly as voltage is increased.
M
INIMUM
C
ONTACT
R
ATING
*
10 mV dc, 10 µA (resistive)
*This value is in reference to a resistive load. Minimum capacity changes depending on switching frequency and environmental conditions
R
ATED
S
WITCH
O
PERATIONS
Mechanical
Electrical
1 x 10
8
(no load)
1 x 10
6
@ 50 V dc, 0.1 A resistive or 10 V dc, 10 mA (resistive)
S
WITCHING
T
IME
< 3 ms
P
ATH
R
ESISTANCE
< 500 mΩ
I
NSULATION
R
ESISTANCE
> 1 x 10
9
Ω
M
AXIMUM
T
HERMAL
O
FFSET
PER
C
HANNEL
(HI-LO)
< 3 μV
C
APACITANCE
Open channel
Channel-mainframe
High-low
< 50 pF
< 20 pF
< 50 pF
B
ANDWIDTH
(-3
dB)
40 MHz (typical)
C
ROSSTALK
(T
YPICAL
)
1 MHz
10 MHz
< -70 dB
< -50 dB
I
SOLATION
(T
YPICAL
)
1 MHz
10 MHz
< -55 dB
< -35 dB
For mating connector, crimp pins, and other accessories, please refer to