© 2019 VPInstruments | MAN-VP-SPRO-EN | Revision: 1901 | Date: 04-07-2019
16
Installation with adjustable safety system
Step 1.
1. Remove the compression fitting's parts on the
long threaded end side
2. Place the safety plate over the compression
fitting long threaded end
Step 2.
1. Mount the nut, and close tight. Then mount
the compression fitting's Teflon rings and nut
Step 3.
1. Keep the ball valve closed!
2. Insert the compression fitting in the ball valve.
Use Teflon tape or liquid sealant.
Step 4.
1. Insert the VPFlowScope probe until it hits the
ball valve. The probe remains in the compression
fitting.
2. Hook the safety line to the ring of the
VPFlowScope probe.