FLOW
DESCRIPTION
FLOW makes the streaming mode data = flow only. If meter is in streaming mode, this
data will stream until a FLOWwSETPOINT, FLOWwSETPOINTwDAC or
TOTALwDIFFwDAC is sent to the meter. If not in streaming mode then one flow will be
return with each FLOW command sent. ReadMeter, WriteMeter both will set flow as the
streaming data.
ASCII STRING COMMANDS
'?Flow + CRC + cr' Version 2.xx read command
'!Flow + value + CRC + cr' Version 2.xx write command
'?Flow + CRC + cr' Version 1.xx read command
Returns ASCII string 'Flow + value + CRC + cr'
Values are a string of digits and an optional decimal place: '10.00'
SETPOINT
DESCRIPTION
SETPOINT sets the control setpoint. The set point is store in the flash memory. This
command is not recommended for version 2.xx,. Use SETPOINT_FLASH.
ASCII STRING COMMANDS
'?Sinv + CRC + cr' Version 2.xx read command
'!Sinv + value + CRC + cr' Version 2.xx write command
'?Sinv + CRC + cr' Version 1.xx read command
'Sinv + value + CRC + cr' Version 1.xx write command
Returns ASCII string 'Sinv + valueSe CRC + cr'
Values are a string of digits and an optional decimal place: '10.00'
CRC=redundancy check bytes; cr=carrage return byte.
SETPOINT_FLASH
DESCRIPTION
SETPOINT_FLASH returns the current flash memory value setpoint with read
command. The write command sets the flash memory value and makes it the active
Summary of Contents for SmartTrak 100 Series
Page 63: ...63 C ADJUSTMENTS Clicking on adjustments brings up the meter settings screen ...
Page 74: ...72 Appendix A C100 M100 SmartTrak Product Specifications Dimensions and Mounting ...
Page 75: ...Vögtlin Instruments Instruction Manual SmartTrak 100 Series 73 ...
Page 76: ...Vögtlin Instruments Instruction Manual SmartTrak 100 Series 4 7 ...
Page 77: ...Vögtlin Instruments Instruction Manual SmartTrak 100 Series 75 ...
Page 78: ...Vögtlin Instruments Instruction Manual SmartTrak 100 Series 76 ...
Page 79: ...Vögtlin Instruments Instruction Manual SmartTrak 100 Series 77 ...
Page 80: ...Vögtlin Instruments Instruction Manual SmartTrak 100 Series 78 ...
Page 82: ...Vögtlin Instruments Instruction Manual SmartTrak 100 Series 80 ...
Page 83: ...Vögtlin Instruments Instruction Manual SmartTrak 100 Series 81 ...
Page 85: ...Vögtlin Instruments Instruction Manual SmartTrak 100 Series 83 ...
Page 86: ...Vögtlin Instruments Instruction Manual SmartTrak 100 Series 84 ...
Page 87: ...Vögtlin Instruments Instruction Manual SmartTrak 100 Series 85 ...
Page 89: ...Vögtlin Instruments Instruction Manual SmartTrak 100 Series 87 ...
Page 90: ...Vögtlin Instruments Instruction Manual SmartTrak 100 Series 88 ...
Page 91: ...Vögtlin Instruments Instruction Manual SmartTrak 100 Series 89 ...
Page 92: ...Vögtlin Instruments Instruction Manual SmartTrak 100 Series 90 ...
Page 109: ...Vögtlin Instruction Manual SmartTrak Series 100 107 private static uint CalcCRC byte cmnd ...
Page 111: ...Vögtlin Instruction Manual SmartTrak Series 100 109 Else ...
Page 112: ...Vögtlin Instruction Manual SmartTrak Series 100 110 crcWord crcWord 2 ...