TIPS - QUESTIONS - ANSWERS
TIPS - QUESTIONS - ANSWERS
5.2
PROBLEMS AND SYSTEM WARNINGS
THE SYSTEM STARTS, BUT THE
CONTROL UNIT DOES NOT
RESPOND
THE SYSTEM DOES NOT START
THE PROBE DOES NOT ARTICU-
LATE PROPERLY
Never squeeze or pull on the articulating
bending neck! This may cause severe da
-
mage to your probe!
A
Check if the main power cable is
connected correctly at both ends and
power is available.
B
Optional battery operation:
Check if the battery pack is fully
charged
C
Check if the battery is properly
connected.
D
Check the systems functionality using
main power.
A
Disconnect and re-connect (hot plug) the
cable to restart and establish the hand-
shake.
B
Is the cable connection from base unit to
control unit established and both ends
connected properly?
A
Check if the micro compressor is turned
ON
B
Check if the micro compressor icon is
displayed in the operation menu
C
Check if the air control meter in the
lower left screen shows sufficient air
pressure (min 6 to 12 bar).
D
Is the optional external air supply turned
on?
E
Is the plug of the external air supply
connector locked?
F
Is there unexpected air flow noise?
G
Check the sealing on the interchange-
able probe connector
H
Is the X-WAY articulation locked?
Press
n
on your control unit.
I
Reel off and straighten the first four me-
ters of your X-WAY probe. Pull the probe
(20 cm before the probe tip) and gently
straighten the insertion tube.