2801/2802 Power Analyzer User Guide
for all interface signals within the interface connectors except for the Ethernet
connections (which are transformer isolated according to the applicable
Ethernet standards). This area is electrically connected to the negative side of
the DC power input via an RF pi filter network having a low DC resistance.
Notes on Grounding:
Between the two ground areas there is a 50ohm (1%) power resistor in
parallel with a 1000pF/50V ceramic capacitor for RF filtering.
The T input connections are isolated from the Internal Electronic Ground
via a nominally 50ohm (1%) power resistor. An isolating current
transducer having sufficient safety rating for the signals being measured
MUST be used when using this input.
The A and Line Switch connections are fully isolated from the Internal
Electronic Ground using transformer isolation to the requirements of
EN61010. This isolation is 100% tested to 4.5kV peak.
The V connections are fully isolated from the Internal Electronic Ground
using transformer isolation to the requirements of IEC61010. This
isolation is 100% tested to 4.5kV peak.
The V connections are resistively isolated from each other with nominally
1Mohm to the requirements of EN61010 in no fault, single fault and
dual fault conditions. This isolation is 100% tested to 2kV peak.
Mechanical isolation is provided between the V connections and the A and
Line Switch connections, which is well in excess of the class ratings of
either set of terminals, there are no electrical signals between these
connections.
Because of the use of transformer isolation, there is very little current
flowing between the UUT terminals and the ground of the product. The
isolation impedance is >10
9
ohms and the capacitance is typically a few
10s of pF.
The EN61010 ratings in each application category for each terminal are
stated on the rear panel of the product next to the terminals.
Summary of Contents for Xitron 2801
Page 1: ...USER GUIDE 2801 2802 Advanced Single and Dual Channel Power Analyzers...
Page 2: ...2 2801 2802 Power Analyzer User Guide...
Page 10: ...10 2801 2802 Power Analyzer User Guide...
Page 101: ...Printing Results 101 Figure 66 Configuration Printout...
Page 102: ...102 2801 2802 Power Analyzer User Guide Figure 67 Waveform Channel 1 Real time...
Page 103: ...Printing Results 103 Figure 68 Waveform Channe1 Distortion...
Page 104: ...104 2801 2802 Power Analyzer User Guide Figure 69 History Chart VOLTS ACDC Trends View...
Page 105: ...Printing Results 105 Figure 70 History Chart AMPS PERCENT THD Extents View...
Page 106: ...106 2801 2802 Power Analyzer User Guide Figure 71 Harmonics Listing page 1...
Page 107: ...Printing Results 107 Figure 72 Harmonics Listing page 2...
Page 145: ...Sending Commands and Interrogatives 145...
Page 147: ...Appendix A Physical Specifications 147...