
Voltage uncertainty is equal to the square root of the selected output
times 0.75
Examples:
Output
Formula
Accuracy
100 VDC
Sqrt(100) * 0.75
± 7.5 VDC
500 VDC
Sqrt(500) * 0.75
± 16.8 VDC
1000 VDC
Sqrt(1000) * 0.75
± 23.7 VDC
5000 VDC
Sqrt(5000) * 0.75
± 53.0 VDC
7000 VDC*
Sqrt(7000) * 0.75
± 62.7 VDC
* Requires option DC-7
DC Voltage Output with option DC-10:
Voltage uncertainty is equal to the square root of the selected output
times 1.5
(AC Voltage specs remain the same as the standard 944 as described above)
Examples:
Output
Formula
Accuracy
100 VDC
Sqrt(100) * 1.5
± 15.0 VDC
500 VDC
Sqrt(500) * 1.5
± 33.5 VDC
1,000 VDC
Sqrt(1000) * 1.5
± 47.4 VDC
5,000 VDC
Sqrt(5000) * 1.5
± 106.1 VDC
7,000 VDC
Sqrt(7000) * 1.5
± 125.5 VDC
10,000 VDC
Sqrt(10000) * 1.5
± 150.0 VDC
DC Current Sensing (DC Leakage)
Range
Resolution
Accuracy
1.800uA
1nA
0.4% of reading ± 6nA
18.00uA
10nA
0.3% of reading ± 30nA
1.800mA
1uA
0.3% of reading ± 30uA
40.0mA
10uA
0.3% of reading ± 300uA
Voltage Coefficient:
15pA/V
DC Current Limits:
Range
Current Limit
100 to 5,000 VDC
35mA
DC Current Limits with option FR-4: