“VitalPatch Off Body” Message
VitalPatch with Adhesive Overlay
If the screen displays the message “VitalPatch Off Body,”
this indicates that the VitalPatch no longer detects
proper skin contact. See figure to the right.
•
If removal was intentional, tap
Disconnect Patch
to
end the monitoring session.
•
If removal was not intentional, inspect the VitalPatch
for adhesive peel-off or lifting on the edges. If the
adhesive looks intact, press the VitalPatch down
against the skin to re-adhere and tap
Wait for
Reconnect
.
If the VitalPatch Reconnects
If reconnection after “VitalPatch Off Body” is successful,
a “Posture Calibration Required” message will be
displayed. Tap
Calibrate
to begin the posture
recalibration process. See
Step 5 and 6 beginning on
.
If you do not recalibrate, VistaPoint will not provide you
with respiration rate or posture while out of calibration.
Accuracy of fall detection and automatic body
temperature is also considered invalid. Heart rate and
live ECG will continue to update continuously.
If the VitalPatch Does Not Reconnect
Replace the VitalPatch if:
•
The VitalPatch does not reconnect after several
minutes.
• The “VitalPatch Off” message continues.
•
The VitalPatch cannot be reapplied by pressing down
to re-adhere.
• The VitalPatch partially or completely falls off the skin.
If the VitalPatch must be replaced, select
Disconnect
Patch
from the
Menu
.
Using the Optional Adhesive Overlay
to Prevent “VitalPatch Off Body”
If the VitalPatch will be exposed to conditions where
lifting and patch-off may occur, you may apply an
optional adhesive overlay to help keep the VitalPatch
adhered to the skin. See
.
Use only the approved adhesive overlay. The use of other
adhesives or tape is not recommended and will prevent
proper operation of the VitalPatch and will void the
warranty of the VitalPatch.
DO NOT
use the adhesive
overlay if the VitalPatch has already lifted off the skin.
The overlay may be used if the VitalPatch begins to peel
up at the edges but has not lost contact with the skin.
Section 19:
“VitalPatch Off Body” Message
Waiting for VitalPatch
to be reattached
VitalPatch b008bf_012345 Off Body
If the VitalPatch is okay, select “Wait
for Reconnect”.
If the VitalPatch has been removed
from the body, or is no longer needed,
select “Disconnect Patch”
Wait for Reconnect
Disconnect Patch
Check that the VitalPatch biosensor is properly adhered to the skin.
Waiting for VitalPatch
Waiting for VitalPatch to be reattached to
the skin.
More Information...
Overlay
VitalPatch