How to check
①
" Movement Inspection" select
②
"Head Contact Sensor Status --> 0 ON" are turned up and head is going down.
Contact sensor is OK if 4090 off turns up when you lift head up by hands.
Go to next step by pressing "select"
③
"Arm Contact Sensor Status --> 0 ON" turns up and chamfer wheel is going down.
Head sensor is OK if 4090 off turns up when you lift chamfer wheel down by the hands.
Go to next step by pressing "select"
④
"Check Encoder Values ---> 0 " turns up and gauge is going down.
Checking the figures by moving gauge to left and right by hands.
It's normal if more than "3000" turns up when moved to the end of left and right.
Go to next step by pressing "select"
⑤
"Did you install a Lens? (Yes->[+], No->[-])" turns up and head goes back
to original position.
That's the MODE to check pressure by inserting pressure guage in to the middle
of head.
Pressure differs from what lens are chosen from first MENU.
When press "No->[-]" and moves to MENU mode.