Dosing system eco-CONTROL EC200-K
Operation and maintenance instructions
Copyright ©
Version 1.8
Maintenance
The Dosing system can be regarded as maintenance-free. Keep the ventilation slots of
the mains adapter and the housing free of dirt. Do not use any aggressive solvents or
cleaners for cleaning, only a damp cloth. Unplug the unit from the mains before doing any
cleaning.
8
Technical data
8.1
Dosing control
Level of medium critical
Not enough medium in the
supply tank
Fill up with the medium. If the
error message remains, check
the sensor and sensor input, if
applicable, brief "emergency
operation" without level
monitoring.
Check the compressed air
Compressed air too little, or
switched off
Check / switch on the
compressed air
Tolerance range set too small
Set the tolerance range to be
larger, see (5.9.6)
Dimensions (HxWxD)
100 x 240 x 260 mm
Weight
approx. 1.3 kg
Voltage
24 V DC, mains adapter supplied as well
Voltage mains adapter
230 V / 50 / 60 Hz
Performance
max. 50 W
depending on the output set
start
Key, foot switch
Start / stop externally
24 V pulse, 10 mA terminal strip; low 0 to 1.5 V,
high 12 to 24 V
On / off switch
yes
Interface
RS 232
Compressed air monitoring
0 to 7 bar (0 to 100 psi)
External memory
MM C/SD memory card min 64 MB max. 24 dosing
programs
can be read with a Microsoft operating system
Operating conditions
+10°C to +40°C (non-condensing), air pressure 1 bar
Medium temperature
+10°C to +40°C
Storage conditions
dry/dust-free -10 to +40°C