Dosing system eco-CONTROL EC200-K
Operation and maintenance instructions
Copyright ©
Version 1.8
Selection whether the calibration is to be done with
quantity unit
ml
(
µl
)
or
g
(mg)
*
and confirm this, the set
quantity
can be set as desired.
* If the saved density value corresponds to the value
1.000 g/cm³ (default value), the input dialog for the
material density appears when you select
g
(mg)
Display 15
• Set and confirm the set quantity.
The value of 0.5 ml preset in the Dosing system is to be regarded as the ideal value.
• Position the calibration vessel underneath the outlet nozzle (dosing needle) of the
dispenser.
• Press the
START
key, the calibration starts, the dispenser is switched on and
conveys the quantity specified by the system (recommended 0.5 ml/g). This process
can be repeated as often as desired. e.g., for comparative measurements.
• The ACTUAL amount that goes into the calibration vessel is determined in the
previously set measuring unit. See also 5.9.1, Dosing amount.
• Input and confirm the ACTUAL quantity that has been determined.
• Select the OK button and confirm, the next dialog to save the calibration value
appears (display 17.1).
• Accept or discard the suggested program slot, confirm
this, and the system has been calibrated.
Displ
ay 17
.1
5.8.2
Calibration manual
The
Calibration manual
function provides a simplified option to recalibrate the system
without having to run through the entire process in the
Calibration auto
menu.
This can be helpful, for example, after changing the stator, in the event of batch-
related product variations, etc., whereby the set dosing value is not to be changed
(QA documents, etc.).
Caution:
Setting the flow rate in the
Calibration manual
menu also changes the
dosing results of all the dosing programs by a linear factor.
Press Start
Esc
Calibration auto
Desired qty
ml
g
OK
Actual qty
Flow rate
00.500 ml
middle
00.000 ml
Save with OK
Esc
Calibration auto
Save calibration value
at prog. no.: 00
OK
OK