D122 REAR PANEL OPERATOR CONTROLS 

(Refer to Figure #2) 

 
10.  PULSE O/C INPUT.  Allows for PULSED OPEN/PULSED CLOSED shutter operation. Sets shutter 
open on first falling edge. Closes shutter on next falling edge.  
 
11.  PULSE OUTPUT.  Active-low output. The duration of the low pulse is equal to the duration of the 
shutter exposure pulse.  This output can be used for daisy chaining      D122 units together. 
 
12.  RESET INPUT. BNC Active-low input. Resets D122 disabled by interrupt (mode A  or mode B) when 
switched to ground by external low level logic or simply shorting to ground through a switch contact. A 
low level input also returns shutter to position determined by the N.O/N.C. switch. 
 
13.  PULSE INPUT. Active-low input. Control of shutter exposure and frequency from a low going signal 
source. The pulse duration determines the shutter exposure. The      frequency of this input pulse will be 
equal to the frequency of shutter exposures.  
 
14.  SYNC. OUTPUT. Active-low output for shutters equipped with Solid State Sync. system. The 
shutter's internal sync. circuit sets a low output at this BNC connector when the shutter's sync. system is 
active. 
 
15.  HIGH/LOW Switch. Pulse energy toggle switch. Switch to HIGH position for operation of VS35 
shutters only. Use STD position for all other 

UNIBLITZ

 shutters. 

      
16.  SHUTTER.  7-Pin female connector to mate with 7-pin male connector of 10' shutter interconnect 
cable provided with the control.  
 
17.  115VAC/230VAC. Line voltage connector. Mates with female plug on line cord provided with D122.   
See SPECIFICATIONS for additional information. 
 
18.  LINE FUSE. Replace with 3AG 1/2 amp slo-blo fuse.  When operating on 230VAC replace line fuse 
with 1/4 amp SB. 
 
19.  SHUTTER FUSE. Replace with 3AG 1/2 amp slo-blo fuse. 
 
20.  RS-232C. DB-9 female connector. Provides access to D122 RS-232C interface, allowing the user to 
control the D122 via an RS-232C serial input port. See INPUT SPECIFICATIONS section for more 
information. 

Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Summary of Contents for UNIBLITZ D122

Page 1: ...service in house repair center WE BUY USED EQUIPMENT Sell your excess underutilized and idle used equipment We also offer credit for buy backs and trade ins www artisantg com WeBuyEquipment REMOTE IN...

Page 2: ...VINCENT ASSOCIATES USER MANUAL Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...

Page 3: ...DRIVER TIMER Vincent Associates 803 Linden Avenue Rochester New York 14625 Service Orders 800 828 6972 In NY State 585 385 5930 Fax 585 385 6004 A Division of VA Inc Version 1 3 1995 Artisan Technolog...

Page 4: ...LS 16 Figure 3 T132 Front Panels Operator Controls 15 T132 REAR PANEL OPERATOR CONTROLS 18 Figure 4 T132 Rear Panel Operator Controls 17 OPERATING INSTRUCTIONS 20 INITIAL OPERATION AND TESTING 20 SYST...

Page 5: ...errupting the exposure cycle One feature of the D122 is the selection of interrupt mode With the mode switch in the A position the unit will disable itself upon a AC line failure When power is restore...

Page 6: ...tion of trigger to pulse applied to driver Typical 5 7 S delay Shutter Drive Continuously variable frequency of exposures from DC to the shutter s maximum rate maximum peak pulse power 400 W pulse vol...

Page 7: ...put pulse to ensure reset 100 s Pulse O C BNC Input impedance 10K ohms Active low Source current 500 A Minimum pulse to activate 300 s Maximum voltage required to ensure operation 5VDC First pulse inp...

Page 8: ...ent 25mA Power supply Control 4 5VDC regulated output T132 only provided for use in remote switching and or control circuits Power supply Control 1 8VDC unregulated output D122 only provided for use i...

Page 9: ...length 7 pin male connector to 7 pin female connector Accessories optional Model 710R Remote Trigger Cable 10 ft length Hand held push button switch to BNC connector Accessories optional Model 710R F...

Page 10: ...TD A POWER N C RESET SHUTTER N O SYNC DISABLED SHUTTER DRIVER MODEL D122 ON OFF Figure 1 D122 Front Panel Operator Controls Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE ww...

Page 11: ...the shutter will return to the position set by an external input or the position of the N O N C switch A AC interrupt mode Prevents the driver from reopening the shutter immediately upon power up if...

Page 12: ...SET Switch Resets D122 disabled by interrupts mode A or mode B and returns shutter to the position determined by the N O N C or the status of the PULSE input The PULSE O C flip flop is also reset If t...

Page 13: ...12 NOTES Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...

Page 14: ...SE INPUT INPUT INPUT PULSE O C SHUTTER HIGH LOW SHUTTER FUSE FUSE LINE 115 VAC 60 Hz 220 VAC 50 Hz RESET Figure 2 D122 Rear Panel Operator Controls Artisan Technology Group Quality Instrumentation Gua...

Page 15: ...er exposures 14 SYNC OUTPUT Active low output for shutters equipped with Solid State Sync system The shutter s internal sync circuit sets a low output at this BNC connector when the shutter s sync sys...

Page 16: ...10 ms 1 sec 100 ms 1 ms 1ms 1ms 1 ms MODEL T132 SHUTTER DRIVER TIMER POST PRE NORM CONT ON OFF ACTUATE N O N C RESET SHUTTER SYNC Figure 3 T132 Front Panels Operator Controls Artisan Technology Group...

Page 17: ...l be timed opened N C or closed N O In the N C position the shutter will be opened for the desired exposure In the N O position the shutter will be timed in the closed position Any RESET command will...

Page 18: ...INPUT TRIG OUTPUT DELAY INPUT RESET SHUTTER HIGH LOW CONTROL SHUTTER FUSE FUSE LINE 115 VAC 60 Hz 220 VAC 50 Hz Figure 4 T132 Rear Panel Operator Controls Artisan Technology Group Quality Instrumenta...

Page 19: ...of your driver Each terminal is explained below CONTROL 1 Active high input TRIGGER The rising edge of a trigger signal starts EXPOSURE and DELAY sequence This input provides access to the T132 timer...

Page 20: ...uration set on the front panel when the shutter DELAY is activated In the PRE mode this occurs upon trigger of T132 In the POST mode this occurs at the end of the EXPOSURE interval 26 RS 232C DB 9 fem...

Page 21: ...SYNC output will go low when the shutter is open The absence of the solid state synchronization option will only inhibit the operation of the SYNC output and LED The remainder of the controller system...

Page 22: ...unit s front panel or through the input on the rear panel will de activate the shutter and clear the RS 232 output latches The following is a test program written to test the input commands to the RS...

Page 23: ...1 CHR J 3 PRINT CONTROL TRIGGER COMMAND SENT GOSUB TIMEOUT GOTO MENU ELSEIF A R THEN PRINT 1 CHR J 4 PRINT CONTROL RESET COMMAND SENT GOSUB TIMEOUT GOTO MENU ELSEIF A Q THEN L 0 PRINT PROGRAM TERMINAT...

Page 24: ...d been opened by a RS 232 command or the PULSE O C input when the unit is reset the RS 232 must send another OPEN command or the PULSE O C must be activated again In the STD mode all functions of the...

Page 25: ...30 0 35 0 40 0 45 0 50 0 55 0 60 0 65 0 Exposure Time msec Pulse Voltage 5 Hz 10 Hz 15 Hz 20 Hz 25 Hz 30 Hz 35 Hz 40 Hz Pulse Voltage vs Frequency at Select Exposure Time 0 0 10 0 20 0 30 0 40 0 50 0...

Page 26: ...25 Figure 7 Overall Housing Dimensions Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...

Page 27: ...ertently adjusted improperly Proper care and maintenance of the unit should be taken as with any electronic instrument WARRANTY Great care has been taken to ensure that our products are free from defe...

Page 28: ...service in house repair center WE BUY USED EQUIPMENT Sell your excess underutilized and idle used equipment We also offer credit for buy backs and trade ins www artisantg com WeBuyEquipment REMOTE IN...

Reviews: