74
8-5-8
Readability Setting
The larger the readability becomes, the less the scale is affected by external influences. In addition, it
takes less time for the scale reading to stabilize.
(1)
(2)
Refer to “Appendix 1-1 Basic specifications” and “Appendix 4 Weighing capacity and readability
by unit” for the default readability of each unit.
This function is not available on HJ62K0.1DS(R).
1
Select the readability setting.
- Press [Menu] key, then press [1]-[4] keys to go
to <64A READABILIT>.
- Press [4] key to change the setting value.
- Press [1]/[2] key to select.
The default readability
1
5
Set value
1:
1(default)
5(default)
2:
2
10
5:
5
20
10:
10
50
- Press [4] key to fix.
2
Exit the setting menu.
- Press [Menu] key to shift to the measuring
mode.
8-5-9
Span adjustment with internal weight at power-on
(1)
(2)
This function is available only on models with internal calibration device.
“Span adjustment with internal weight at power-on” operates first power-on after connection of
AC adapter.
1
Select the span adjustment at power-on.
- Press [Menu] key, then press [1]-[4] keys to go
to <64B START CAL>.
- Press [4] key to change the setting value.
- Press [1]/[2] key to select.
OFF: Disable
FORCE: Enable
Always when the power is ON.
SELEC: Selectable
Message is displayed.
- Press [4] key to fix.
2
Exit the setting menu.
- Press [Menu] key to shift to the measuring
mode.
3
Operate span adjustment at power-on.
- Press [On/Off] key to turn on the scale.
- When <SELEC> is selected at step 1, select
whether or not execute span adjustment by
pressing [3]/[4] key.
YES: Execute
NO: Not execute
The scale executes span adjustment by internal
weight and then shifts to measuring mode.
Do not power-off the scale while span adjustment is operating.