19 

2-4

 

Weighing the additional sample 

Weigh the first sample and the additional sample separately. 

 

Place a sample to be weighed. 

 

 

 

 

 

 

The mass of the sample to be 
weighed is indicated.

 

Perform tare subtraction. 

 

 

 

 

 

 

 

 

Press [Tare] key. 

The indication changes to zero and 
the < 

  > symbol appears. 

Place an additional sample to be weighed. 

 

 

 

 

 

 

The mass of the added sample 
alone is indicated. 

 

2-5

 

Basic operation 

 

 

 

Shortcuts for various modes/functions can be assigned to functional keys. Please refer to   
“8-2 Shortcut setting for accessing various measuring modes” and “8-3 Free key setting”. 

 

2-5-1

 

Hierarchy of a setting menu 

The setting menu of this product is divided into four, from the first layer to the third layer and for various 
settings. 

 

   

 

Summary of Contents for HJ-K Series

Page 1: ...cale Series Operation Manual SHINKO DENSHI CO LTD IMPORTANT To ensure safe and proper use of the scale please read this manual carefully After reading this manual store it in a safe place near the sca...

Page 2: ...otice This document has been created carefully In case that however any error or imperfection is found by any chance please let us know Documents of which pages are missing or irregularly bound will b...

Page 3: ...hese dangers In addition there are so many impossible and don ts and so writing all of them in the operation manual is impossible Therefore it is safe to think that what is not written in the operatio...

Page 4: ...utions are not taken Used for accurate weighing and appropriate usage of the equipment Used for referenced information which is useful for product operation Used for Prohibition items Used for Mandato...

Page 5: ...How to interpret the display 15 1 8 1 Description of segment 15 1 8 2 LCD character font 16 2 Basic usage 17 2 1 Turning on off the power and checking for the operation 17 2 2 Zero point adjustment 1...

Page 6: ...t functions 47 6 2 Connector terminal numbers their functions and specifications 48 6 2 1 D SUB9P Connector for RS232C I O 48 6 2 2 D SUB9P Connector for serial output for peripherals 49 6 3 Communica...

Page 7: ...sult 72 8 5 4 Date indication format 72 8 5 5 Date setting 72 8 5 6 Time setting 73 8 5 7 Output language 73 8 5 8 Readability Setting 74 8 5 9 Span adjustment with internal weight at power on 74 8 5...

Page 8: ...y cause the sample to fall from the weighing pan leading to a bodily injury or destruction of the sample Do not route the cables across passages The cables could be tripped on by a passed by and the s...

Page 9: ...samples on the scale Do not use volatile solvents for cleaning anything other than the weighing pan The key panel dust waterproof packing or other resin components of could deform and the scale no lo...

Page 10: ...posed to air current from an air conditioner or to direct sun generate abrupt temperature change and resultantly cause unstable weight indication and therefore should be avoided Gravity The latitude a...

Page 11: ...autions A dust cover if equipped for the scale may possibly make the weight indication unstable due to static electricity charged on the cover at a low humidity When this is the case wipe the cover wi...

Page 12: ...ontains the following If anything missing or broken should be found please inform the store where you purchased the product Weighing unit and indicator set 1 Cable cover 1 Base cover 1 Indicator mount...

Page 13: ...emovable 8 RS 232C connector D sub 9 pin male Mount the connector cover when not connected 9 Connector for peripheral devices D sub 9 pin male Mount the connector cover when not connected 10 Cover of...

Page 14: ...without integrating the indicator unit 1 Remove the weighing pan of the weighing unit 2 Slightly open both sides of the cable cover and mount it vertically in the cable housing Then mount the base co...

Page 15: ...icator mounting bracket and mount it vertically in the cable housing Check that the supporting boards are hooked on the bottom of the cable housing Then fasten the knobs 3 Fold the cable and put it in...

Page 16: ...9 5 Put back the weighing pan 6 Then insert the indicator unit in the indicator mounting bracket at an angle 7 Mount the base cover vertically...

Page 17: ...ing wet When not connected be sure not leave covers of the AC adapter jack and Dsub 9P connectors open 1 Remove the covers for D sub 9P connectors and connect RS232C cables when necessary 2 Put the AC...

Page 18: ...the direction shown in the figure on the left to loosen them 2 Level the scale 1 While watching the level turn the adjusters provided on the bottom to level the main unit 2 Bring the bubble enters in...

Page 19: ...ording to the following manual 1 Unplug AC adapter and interface cables 2 When the separate type bundle the scale cable not to hang on the floor 3 Holding position to lift up and carry Bottom surface...

Page 20: ...em Use for moving down to the menu item selections or use for decrementing the numeric value 8 3 Use for selecting the mode function and item Use for moving to the upper menu layer or use for selectin...

Page 21: ...Zero Use for changing polarity 3 Tare Input a decimal point in the Multiplied by Coefficient mode 4 1 Use for incrementing the numeric values 0 1 2 9 0 5 2 Use for decrementing the numeric values 0 9...

Page 22: ...preset tare weight is being subtracted 9 Preset tared weight Indicates the preset tare weight is being subtracted 10 gram Indicates the gram unit 11 kilogram Indicates the kilogram unit 12 16 segment...

Page 23: ...R S T U V W X Y Z c comma point 1 2 3 4 5 6 7 8 9 0 space minus hyphen 16 segment A B C D E F G H I J K L M N O P Q R S T U V W X Y Z b c d g l m n o t w 1 2 3 4 5 6 7 8 9 0 asterisk slash left arrow...

Page 24: ...e Press and hold On Off key About 2 seconds 1 Pressing and holding On Off key obtains the standby status from any operation status 2 The scale starts up in the last measuring mode before it was switch...

Page 25: ...5 Clear the tare weight data Remove the sample and tare on the weighing pan then press Zero key Therefore the indication becomes zero and indication disappears 1 Performing the tare narrows the weigh...

Page 26: ...the symbol appears 3 Place an additional sample to be weighed The mass of the added sample alone is indicated 2 5 Basic operation Shortcuts for various modes functions can be assigned to functional ke...

Page 27: ...Operation of the setting menu To perform settings for various functions from the state of weighing chiefly execute the following procedure Go to the menu item to set Select the setting value and exec...

Page 28: ...o change the polarity to 2 Input 1 The digit for inputting is blinking Press 1 2 key to increment decrement the digit to 1 Press 3 key to input the next digit 3 Input 2 3 4 5 Input 2 3 4 5 by the proc...

Page 29: ...key F2 and F5 2 key F3 and F6 3 key Please take care not to confuse F1 F4 to 1 4 keys 2 Refer to 8 Controlling and adjustment functions for assigning Free keys and Shortcuts to functional keys Weighi...

Page 30: ...Kong tael TLS Singapore Malaysia tael TLT Taiwan tael TOLA tola Comparator function 13 COMPARATOR ON OFF 131 ACTIVATE OFF OFF H L Upper and lower limits HIGH Upper limit alone valid LOW Lower limit a...

Page 31: ...id Back light setting 1A BACKLIGHT OFF OFF 3MIN 3 minutes 5MIN 5 minutes 10MIN 10 minutes 30MIN 30minutes ON Always ON Auto power off 1B AUTO OFF OFF Invalid 3MIN 3 minutes 5MIN 5 minutes 10MIN 10 min...

Page 32: ...xecuted then the indication changes to zero and the symbol lights 4 Weigh the sample Place the sample The weighing result is displayed 3 2 2 Counting mode This mode is not legal for trade Counting mod...

Page 33: ...ghing pan Press Tare key Place the set number of samples on the weighing pan 4 Record the unit weight Press 4 key to fix The unit weight is recorded 5 Simple SCS method When enabled When 1C Simple SCS...

Page 34: ...x 99 Add blinks on the display and unit weight cannot be updated In this case add samples until Add indication disappears or select the larger number of samples in step 2 3 When simple SCS is operati...

Page 35: ...ected go to step 4 2 Select the unit weight value input mode Press 1 2 key to select PCSWGT Unit weight value input Press 4 key to fix 3 Input the unit weight Numeric value input Refer to 2 5 3 Numeri...

Page 36: ...key then press 1 4 keys to go to 11 MODE Press 4 key to change the setting value Press 1 2 key to select PCNT Percentage mode Press 4 key to fix 2 Exit the setting menu Press Menu key to shift to the...

Page 37: ...30 3 3 1 Switching the display at percentage mode 1 Press 1 4 keys to switch the display ADD and TOTL can be used when the 14 ADDITION is activated...

Page 38: ...to fix 2 Exit the setting menu Press Menu key to shift to the Multiplied by Coefficient mode 3 Select whether or not employ the previous recorded coefficient Press 3 4 key to select whether or not emp...

Page 39: ...it Only g kg and ct are available for verified scale 1 Select the unit setting Press Menu key then press 1 4 keys to go to 12 UNIT Press 4 key to change the setting value Press 1 2 key to select the u...

Page 40: ...etting lower limit Single point setting upper limit Two point setting upper and lower limits Over the upper limit Blinking Blinking Blinking Appropriate amount Blinking Blinking Blinking Below the low...

Page 41: ...s 1 2 key to select FULL At all times STBL Only at stable times Press 4 key to fix 3 Select the discriminant range Press 1 2 keys to go to 133 RANGE Press 4 key to change the setting value Press 1 2 k...

Page 42: ...nction Method of weighing samples to be weighed without replacing the samples Net adding function The adding function can be used in Weighing mode Percentage mode Counting mode and Multiplied by Coeff...

Page 43: ...le to be weighed After appears press 1 ADD key The weighed value is stored and SUM TOTAL is indicated for a few seconds Repeat this operation to perform addition In the case of the net addition Add a...

Page 44: ...ph 1 Select the bar graph indication Press Menu key then press 1 4 keys to go to 18 BARGRAPH Press 4 key to change the setting value Press 1 2 key to select OFF Invalid ON valid Press 4 key to fix 2 E...

Page 45: ...e following conditions 1 Setting menu is being displayed 2 A sample is placed on the weighing pan and the display is not stable When is not displayed 3 12 Buzzer setting Setting buzzer Buzzer beeps wh...

Page 46: ...ress Menu key to shift to the measuring mode 3 14 Range mode setting This function is to enable disable automatic readability d switching of HJ62K0 1DS R double range model Single range mode Double ra...

Page 47: ...ration mode Automatic Zero Tracking 23 ZERO TRAC OFF OFF 0 5 0 5 d s 1 1 d s 2 2 d s 4 4 d s 4 2 Stability discrimination width When the larger numeric value is set in this setting menu the laxer stab...

Page 48: ...nu key to shift to the measuring mode 4 4 Zero tracking Setting to the zero tracking function makes it possible to automatically correct the zero point fluctuation caused by the temperature fluctuatio...

Page 49: ...Setting value input Weight Comparator 33 COMPARE WEIGHT Upper limit value setting 331 WEIGHT HIGH Setting value input Target value setting 332 WEIGHT REF Setting value input Lower limit value setting...

Page 50: ...xecute 4 Setting 4 execute 5 Setting 5 execute 2 Exit the setting menu Press Menu key to shift to the measuring mode Preset tared weight is displayed with indication when preset tare weight value is a...

Page 51: ...e 5 2 2 2 Numeric value setting method 1 Set a preset tare weight value Numeric value input Refer to 2 5 3 Numeric value input Press 4 key to select NUM Numeric value Input the preset tare value Press...

Page 52: ...values that have been set for the reference numeric value For example Two point upper and lower limits setting Reference value 1000 0g Lower limit value 900 0 g Upper limit value 1200 0 g Discriminat...

Page 53: ...is equivalent to the upper limit value Press 4 key to fix The upper limit value is recorded 5 3 2 Numeric value setting method 1 Set a lower limit value Numeric value input Refer to 2 5 3 Numeric valu...

Page 54: ...412 FORMAT 6 6 digit numeric format 7 7 digit numeric format 8 8 digit numeric format CSP6 CSP 6 digit format CSP7 CSP 7 digit format 1 CBM CBM format Output condition 413 CONDITION OFF Output stop 1...

Page 55: ...High level data logic 0 5 to 15 V Low level data logic 1 5 to 15 V Baud rate 1200 2400 4800 9600 19200 38400 57600 115200 bps Transmission code Composition Start bit 1 bit Parity bit None Odd number E...

Page 56: ...cale 1 Data composition Measurement result 6 digit numeric format CSP 6 digit format Consists of 15 characters including terminators CR 0x0D LF 0x0A 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 P1 D1 D2 D3 D4...

Page 57: ...nce L B 0x4C 0x42 lb pound O T 0x4F 0x54 ozt troy ounce D W 0x44 0x57 dwt pennyweight G R 0x47 0x52 GN grain T L 0x54 0x4C tlH Hong Kong tael T L 0x54 0x4C tlS Singapore Malaysia tael T L 0x54 0x4C tl...

Page 58: ...20 0x20 Net weight 41A STATUS ON P T SP SP SP SP 0x50 0x54 0x20 0x20 0x20 0x20 Preset tare weight T SP SP SP SP SP 0x54 0x20 0x20 0x20 0x20 0x20 Tare weight T O T A L SP 0x54 0x4F 0x54 0x41 0x4C 0x20...

Page 59: ...l completion or if the command is invalid an error the scale will transmit an error response When the scale is in normal display mode it usually sends a response to a command within one second of rece...

Page 60: ...le times Continuous output at unstable times O 7 0x4f 0x37 Press down Output key for one time output at stable times O 8 0x4f 0x38 One time instant output O 9 0x4f 0x39 One time output after stability...

Page 61: ...input while the other mode operation the scale output an abnormal response 4 If the input value is invalid the scale output an abnormal response 6 4 3 1 Comparator setting command C1 C2 Code C1 Code C...

Page 62: ...nd A1 code A1 Description ACK 0 06 Normal response NAK 0 15 Abnormal response 6 6 Communication setting For verified scale Setting menu 412 FORMAT is not available That is fixed to CBM CBM format and...

Page 63: ...ry time when the scale reaches stable Output stop at unstable times 6 Continuous output at unstable times and one time output every time when the scale reaches stable 7 One time output after Output ke...

Page 64: ...ERATION OFF Locking disabled ON Setting menu is hidden and unmodifiable Functions relating to the performance 532 PERFORM OFF Locking disabled ON Setting menu is hidden and unmodifiable Preset tare an...

Page 65: ...k function Various setting menus can be locked 1 Select the menu lock function Press Menu key then press 1 4 keys to go to 53 MENU LOCK Press 4 key to change Press 1 2 key to select Refer to Set List...

Page 66: ...without internal calibration weight ICAL for model with internal calibration weight Initial setting value Functions relating to the lock 6 ADMIN ADJUST Key assignment for mode selection 61 SHORTCUT MO...

Page 67: ...ord registration 644 SET USER PASS Setting value input Span adjustment result output 645 SPAN OUT OFF OFF ON Valid Date format 646 DATE DISP Y M D yyyy mm dd D M Y dd mm yyyy M D Y mm dd yyyy Date set...

Page 68: ...o go to 611 F1 KEY Press 4 key to change Press 1 2 key to select 611 F1 KEY F1 above 1 key 612 F2 KEY F2 above 2 key 613 F3 KEY F3 above 3 key 2 Select the measuring modes Press 4 key to change the se...

Page 69: ...EY F2 623 F3 KEY F3 624 F4 KEY F4 625 F5 KEY F5 626 F6 KEY F6 2 Select the function to assign to the Free key Press 4 key to change the setting value Press 1 2 key to select Refer to Set List Press 4...

Page 70: ...span adjustment significantly affects the weighing accuracy Please read this procedure carefully before getting to the adjustment 8 4 1 1 Span adjustment with external weight This function is unavail...

Page 71: ...ss Menu key to shift to the measuring mode 1 List of weights used for the span adjustment by model Unit g Model HJ17K0 1S R HJ22K0 1S R HJ33K0 1S R HJ62K0 1DS R Selectable weight on the menu 17000 220...

Page 72: ...For HJ17K0 1S R and HJ22K0 1S R For HJ33K0 1S R and HJ62K0 1DS R Place the weight in the centre of the weighing pan For HJ17K0 1S R and HJ22K0 1S R Display changes to the order of on FS blinking of on...

Page 73: ...Y blinking of Ch 0 Ch FS blinking of Ch FS 633 INT CAL 3 Exit the setting menu Press Menu key to shift to the measuring mode 8 4 1 4 Span test with internal weight Do not power off the scale while thi...

Page 74: ...Please read this procedure carefully before getting to the adjustment 4 Do not power off the scale while this function is operating This function is available only on models with internal calibration...

Page 75: ...from the weighing pan Display automatically changes to blinking of on 0 and zero point adjustment starts On completion of the zero point adjustment the display automatically changes to blinking of Ch...

Page 76: ...generates an alarm when calibration is needed This function is unavailable for verified scale 1 Select the Advice CAL Press Menu key then press 1 4 keys to go to 638 Advice CAL and then press 4 key to...

Page 77: ...re above Press 4 key to fix the scale ID and shift to 641 SCALE ID 3 Exit the setting menu Press Menu key to shift to the measuring mode 8 5 2 Password control Enable disable the password protection 1...

Page 78: ...y to shift to the measuring mode 8 5 2 2 User password registration Administrator can register the user password for each user operator 1 Refer to Appendix 5 Scale operation with password control func...

Page 79: ...go to 646 DATE DISP Press 4 key to change the setting value Press 1 2 key to select Y M D Year Month Day D M Y Day Month Year M D Y Month Day Year Press 4 key to fix 2 Exit the setting menu Press Menu...

Page 80: ...hen TIME is assigned to 3 key Press 1 3 key of which the free key TIME is assigned The time is indicated Press 4 key to return to the measuring mode 8 5 7 Output language Output language can be select...

Page 81: ...ustment with internal weight at power on 1 2 This function is available only on models with internal calibration device Span adjustment with internal weight at power on operates first power on after c...

Page 82: ...tion due to an unexpected power failure in a place where the power supply is unstable This function is not available on verified scale 1 Unlike its name Save Tare Weight Function or its displayed name...

Page 83: ...ment the date and time setting 1 Select the initialise Press Menu key then press 1 4 keys to go to 64E INITIALIZE Press 4 key Press 3 4 key to select NO Cancel YES Execute After initialisation is comp...

Page 84: ...Initial zero adjustment error The initial zero adjustment was not completed in the process of starting from the standby status because of the unstable load Improper setting of the weighing pan or pan...

Page 85: ...7 Time out error of importing the sample weight in the actual value setting method at Comparator function Improper setting of the weighing pan or pan base is suspected Check for contact with other obj...

Page 86: ...off disconnect the AC adapter and close the AC adapter jack with the rubber cap disconnect RS232 cables and close the D sub9p connector with cover 2 Remove the weighing pan 3 Cleaning method 1 Wipe d...

Page 87: ...ernal HJ22K0 1S External 22000 g 22 kg 110000 ct 0 1 g 0 0001 kg 0 5 ct 0 1 g 0 0001 kg 0 5 ct 10 g 0 01 kg 50 ct HJ22K0 1SR Internal and External HJ33K0 1S External 33000 g 33 kg 165000 ct 0 1 g 0 00...

Page 88: ...5 g 0 005 kg 25 ct 1 g 0 001 kg 5 ct 0 1 g 0 0001 kg 5 ct II 0 1 g 0 0001 kg 5 ct 10 g 0 01 kg 500 ct HJ22K0 1SR Internal and External HJ33K0 1S External 33000 g 33 kg 165000 ct 5 g 0 005 kg 25 ct 1...

Page 89: ...t Date Time setting Output language English Japanese Readability Span adjustment at power on Direct start Initialise Other functions which can be assigned to free keys Gross indication Tare value indi...

Page 90: ...Humidity 80 RH or lower no condensation Pollution degree 2 Altitude 2000 m or less above sea level Location of use Indoor use only Ingress Protection IP65 Option Relay output factory option 1 RS422 ou...

Page 91: ...84 Appendix 2 Dimensional outline drawing...

Page 92: ...at 5 00000000E 00 1 pound 2 2046226E 03 1 ounce 3 5273961E 02 1 troy ounce 3 2150746E 02 1 grain 1 5432358E 01 1 pennyweight 6 4301493E 01 1 momme 2 6666667E 01 1 mesghal 2 16999761E 01 1 Hong Kong ta...

Page 93: ...5 0 01 0 1 troy ounce 540 700 1000 Gross 190 09 1900 0 005 0 005 0 005 0 01 0 1 pennyweight 10000 14000 21000 Gross 3900 9 39000 0 1 0 1 0 1 0 1 1 momme 4500 5800 8800 Gross 1600 9 16000 0 05 0 05 0 0...

Page 94: ...tor login mode is indicated on the 7 segment display 3 Select the user to set the authority Select the user 0 Guest user 1 User 1 2 User 2 USER Shift to the User login mode 4 Input the administrator p...

Page 95: ...cessing to setting menus Appendix 5 2 User guest login 1 Power on the scale and go to the User login mode Self check Press On Off key then the scale shifts to User login mode after start up operation...

Page 96: ...rate the scale on the AC adapter When changing to mark you cannot operate Span adjustment with internal weight Span test with internal weight and Calibration of the internal weight When you use those...

Reviews: