Chapter 6
Using LTE/LTE-A TDD Analyzer
Conducting RF measurement
SPA06MA-O User’s Guide
April 2021
22145360, R005
Page 137
Conducting RF measurement
The following sections describe how to conduct RF analysis.
Channel power
The Channel Power measurement is a common test used in the wireless industry to
measure the total transmitted power of a radio within a defined frequency channel. It
acquires a number of points representing the input signal in the time domain, trans
-
forms this information into the frequency domain using Fast Fourier Transform (FFT),
and then calculates the channel power. The effective resolution bandwidth of the
frequency domain trace is proportional to the number of points acquired for the FFT.
The channel power measurement identifies the total RF power and power spectral
density (PSD) of the signal in the LTE channel bandwidth.
Setting measure setup
After configuring the spectrum measurement in
“Configuring spectrum measurements”
, you can set the measure setup to continue your measurement.
To set measure setup
1
Tap the
Setup
(
) icon in the side bar.
2
Tap
Bandwidth
and select the value from the following choices:
1.4 MHz
,
3 MHz
,
5 MHz
,
10 MHz
,
15 MHz
or
20 MHz
.
You can also tap the rectangle with value,(
) under
the chart screen and input the value using the on-screen keyboard.
3
Tap the
Center Frequency
(
)icon under the chart
screen to set the center frequency using the on-screen keyboard.
4
Tap
Subframe Number
and input the value from 0 to 9 using the on-screen
keyboard.
5
Tap to switch
Physical Cell ID
to
Manual
or
Auto
.
–
Auto
lets the instrument detect the Physical Cell ID for the LTE signal auto
-
matically
–
Manual
sets a specific Physical Cell ID for the LTE signal manually in order to
speed up the synchronization with a BTS.
6
If you set the Physical Cell ID to Manual, tap
Physical Cell ID
and input the value
using the on-screen keyboard.
7
Tap
Menu
>
BW/AVG
>
Average
to set the number of measurements to be aver
-
aged using the on-screen keyboard.The input value range is from 1 to 100.
To set limit
1
Tap
Menu > Limit
.
Summary of Contents for SPA06MA
Page 1: ...SPA06MA O User s Guide R005 ...
Page 2: ......
Page 22: ...Chapter 1 Overview Module installation SPA06MA O User s Guide Page 6 22145360 R005 April 2021 ...
Page 410: ...Appendix A Physical specifications SPA06MA O User s Guide Page 394 22145360 R005 April 2021 ...
Page 412: ...SPA06MA O User s Guide 22145360 R005 April 2021 English ...