MPOLx-84/85 (P)
107
16 S
PECIFICATIONS
L
IGHT
S
OURCE
Light Source
Patch cord microscope
Multi-mode
Single-mode
Optical interface / adapter
MTP-PC
MTP-APC
Source type
LED source
Fabry-Perot
laser diode
Wavelengths
850 / 1300 nm
1)
1) Typ/- 20 nm
1310 / 1550 nm
1)
Spectral width (FWHM)
< 50 nm (850 nm) /
< 170 nm (1300 nm)
< 5 / < 5 nm
Spectral width (RMS)
22 / 60 nm
< 5 / < 5 nm
O
utput power
(typical)
-26 dBm
-6
dBm
Launch condition
Encircled Flux
compliant to TIA-
526-14 and IEC
61280-4-1
2)
2) At the output of the frontend connector. Variations between EF measurement
equipment may occur but EF compliance can be expected with a 95 %
confidence factor.
Valid for IEC 61280-4-1 at 850 nm.
Source mode
CW
Stability 15 min
3)
4)
3) Single channel, b5 °C to +45 °C with ΔT = ±0.3 K
after a 20 minutes warm-up
4) In AC operating mode without rechargeable battery, or rechargeable battery
only.
±0.05 dB
Stability 8 h
3) 4)
±0.25 dB
Laser safety
IEC 60825-1:2009
Laser classification
CLASS 1 LASER PRODUCT
Optical interface
FMAE MPO
(many other adapters
available)
Auto pass/fail analysis standards
IEC 61300-3-35 and custom limits
Live image
320 x 240 x 8 bit grey, 10 fps
Light source
Blue LED, 100.000+ hours life
Lighting technique
Coaxial
Magnification field-of-view
low/high
Horizontal: 740/370 μm
Vertical: 550/275 μm
External USB connected P5000i Digital Probe supported.