Remote Commands
SCPI Commands Command Tables
MAP-200 Optical Switch mOSW-C1 Installation and User Guide
Standard
June 2016
Document Number 22112369-120 R000
Page 50
*ESE
Description
Sets the bits in the Standard Event Status Enable register which will determine which events
are reflected in the event summary bit (ESB) of the standard status register. The numeric
value is converted to a binary number. The bits of the register are set to match the bit
values of the binary number.
If bits are set TRUE in the standard event status enable register and a corresponding bit is
set TRUE is the standard event status register, the ESB bit (bit 5) of the status byte register
will be set TRUE.
Input Format
*ESE <Register Value>
Parameter
Type
Value
Register
Numeric
0 <= value <= 255
Standard Event Status Enable Register
Bit 7
Bit 6
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0
Power on
(PON)
User
request
(URQ)
Comman
d error
(CME)
Execution
error
(EXE)
Device
dependent
error
(DDE)
Query
error
(QYE)
Reque
st
control
(RQC)
Operation
complete
(OPC)
Bit 0: operation complete
Bit 1: request control – not supported
Bit 2: query error. It is set when a query error occurs, for example, an attempt is made to read the output queue when the output
queue is empty or when the data in the output queue is lost
Bit 3: device dependent error – not supported
Bit 4: execution error
Bit 5: command error. it is set when a command error is detected by the system, for example, if a syntax error is detected in a
program message, an incorrect command header is received, or an IEEE GET message is received in the middle of a program
message.
Bit 6: user request – not supported
Bit 7: power on
Output Format
None.
Default
0 on power cycle.
Example
The following example sets standard event status enable register bits to 00100000,
indicating that the COMMAND ERROR (CME) byte of the event status register
Terminal Input: