The Blood Pressure Monitor is intended for use in the electromagnetic environment specified
below. The customer or the user of the Blood Pressure Monitor should assure that it is used in
such an environment.
Immunity test
IEC60601 test level
Compliance
level
Electromagnetic
environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±
8kV contact
±
2kV,
±
4kV,
±
8kV,
±
15kV
air
±
8kV contact
±
2kV,
±
4kV,
±
8kV,
±
15kV
air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Electrical fast
transient/ burst
IEC 61000-4-4
±
2kV for power
supply lines
100kHz repetition
frequency
±
1kV for input/output
lines
N/A
--
Surge
IEC 61000-4-5
±
0.5kV,
±
1kV differential
mode line-line
N/A
--
Voltage dips, short
Interruptions and
Voltage variations
on power supply
input lines
IEC 61000-4-11
0% U
T
(100% dip in U
T
)
for 0.5 cycle at 0
°
, 45
°
,
90
°
, 135
°
,180
°
, 225
°
,
270
°
, and 315
°
0% U
T
(100% dip in U
T
)
for 1 cycle at 0
°
70% U
T
(30% dip in U
T
)
for 25/30 cycles at 0
°
0% U
T
(100% dip in U
T
)
for 250/300 cycles at 0
°
N/A
--
Power frequency
(50/60 HZ)
magnetic field IEC
61000-4-8
30 A/m, 50/60Hz
30 A/m,
50/60Hz
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
Note: U
T
is the AC mains voltage prior to application of the test level.
Guidance and Declaration - Electromagnetic Immunity
The Blood Pressure Monitor is intended for use in the specified electromagnetic environment.
The customer or the user of the Blood Pressure Monitor should assure that it is used in such an
environment as described below.
Immunity test
IEC60601 test level
Compliance
level
Electromagnetic
environment - guidance
Conduced RF
IEC61000-4-6
3Vrms
150kHz to 80 MHz
N/A
Portable and mobile RF
communications equipment