background image

3

themselves.

Do not store the product in the following locations: locations in which the product is exposed to
direct sunlight, high temperatures or levels of moisture, or heavy contamination; locations near
to sources of water or fire; or locations that are subject to strong electromagnetic influences.

This product displays changes in the heart rhythm and blood pressure etc. which may have
various different causes. These may be harmless, but may also be triggered by illnesses or
diseases of differing degree of severity. Please consult a medical specialist if you believe you may
have an illness or disease.

Vital signs measurements, such as those taken with this product, cannot identify all diseases.
Regardless of the measurement taken using this product, you should consult your doctor
immediately if you experience symptoms that could indicate acute disease.

Do not self-diagnose or self-medicate on the basis of this product without consulting your doctor.
In particular, do not start taking any new medication or change the type and/or dosage of any
existing medication without prior approval.

This product is not a substitute for a medical examination or your heart or other organ function,
or for medical electrocardiogram recordings, which require more complex measurements.

We recommend that you record the ECG curves and other measurements and provide them to
your doctor if required.

Clean the product and cuff with a dry, soft cloth or a cloth dampened with water and a neutral
detergent. Never use alcohol, benzene, thinner or other harsh chemicals to clean the product or
cuff.

Avoid tightly folding the cuff or storing the hose tightly twisted for long periods, as such
treatment may shorten the life of the components.

The product and cuff are not water-resistant. Prevent rain, sweat and water from soiling the
product and cuff.

To measure blood pressure, the arm must be squeezed by the cuff hard enough to temporarily
stop blood flow through the artery. This may cause pain, numbness or a temporary red mark to
the arm. This condition will appear especially when measurement is repeated successively. Any
pain, numbness, or red marks will disappear with time.

Too frequent measurements can cause injury to the patient due to blood flow interference.

Consult your physician before using this product on an arm with an arterio-venous (A-V) shunt.

Consult your physician before using this monitor if you have had a mastectomy or lymph node
clearance.

The pressurization of the CUFF can temporarily cause loss of function of simultaneously used
monitoring product on the same limb.

Consult your physician before using the product if you have severe blood flow problems or blood
disorders as cuff inflation can cause bruising.

Please prevent that operation of the product results in prolonged impairment of the circulation

Summary of Contents for BP2

Page 1: ...1 User s Manual Blood Pressure Monitor Model BP2 BP2A ...

Page 2: ...heir health We recommend not to use this product if you have a pacemaker or other implanted products Follow the advice given by your doctor if applicable Do not use this product with a defibrillator Never submerge the product in water or other liquids Do not clean the product with acetone or other volatile solutions Do not drop this product or subject it to strong impact Do not place this product ...

Page 3: ... the ECG curves and other measurements and provide them to your doctor if required Clean the product and cuff with a dry soft cloth or a cloth dampened with water and a neutral detergent Never use alcohol benzene thinner or other harsh chemicals to clean the product or cuff Avoid tightly folding the cuff or storing the hose tightly twisted for long periods as such treatment may shorten the life of...

Page 4: ...by inadequately trained personnel may result in damage or burn The patient is an intended operator Do not carry out the servicing and maintenance while the product is in use The patient can safely use all the functions of the product and the patient can maintain the product by carefully reading Chapter 7 This product emits radio frequencies RF in the 2 4 GHz band DO NOT use this product in locatio...

Page 5: ... device is a blood pressure monitor intended for use in measuring blood pressure and pulse rate in adult population The product is intended to measure display store and review adults single channel ECG rhythms and gives some suggested symptoms such as regular beat irregular beat low HR and high HR 2 2 Contraindications This product is contraindicated for use in ambulatory environments This product...

Page 6: ...s Symbol Meaning SYS Systolic blood pressure DIA Diastolic blood pressure Pulse beat HR Heart rate Battery indicator Charging Application part type BF Manufacturer IP22 Against ingress of solid foreign objects 12 5mm diameter Against dripping 15 tilted Follow operating instructions MRI unsafe Presents hazards in all MR environments as product contains strongly ferromagnetic materials Non ionizing ...

Page 7: ...elaxed and comfortably seated Sit in a chair with your legs uncrossed and your feet flat on the floor Place your left arm on a table so the cuff is level with your heart Note The blood pressure can differ between the right arm and the left arm and the measured blood pressure readings can be different The Viatom recommends to always use the same arm for measurement If the blood pressure readings be...

Page 8: ...a in time NIBP Measurement Principle The NIBP measurement way is oscillation method Oscillation measurement is using automatic inflator pump When the pressure is high enough to block arterial blood flow then it would deflate slowly and record all the change of cuff pressure in the deflation process to calculate blood pressure based on certain algorithm The computer will judge whether the quality o...

Page 9: ...o not touch your body with the hand with which you are taking the measurement Please note that there must be no skin contact between your right and left hand Otherwise the measurement cannot be taken correctly Stay still during the measurement do not speak and hold the product still Movements of any kind will falsify the measurements If possible take the measurement when sitting and not when stand...

Page 10: ... The heart rate 50 min If the measurement results do not meet the Irregular beat High HR and Low HR then judge the Regular beat 3 4 Bluetooth The product Bluetooth will be enabled automatically only when the screen lights up 1 Ensure the product screen is on to keep the product Bluetooth enabled 2 Make sure the phone Bluetooth is enabled 3 Select the product ID from the phone then the product will...

Page 11: ... pairing process away from possible interference source The product doesn t response to the button press The battery may be low Charge battery and try again The product is running in an unexpected status Reset the product by press and hold the button for 8s The device might be damaged Please contact your local distributor Cannot get blood pressure measured results The measurement is interrupted by...

Page 12: ...ance IP22 Drop test 1 0 m Physical Size 135 mm 46mm 21mm main unit Weight about 220 g main unit Cuff size 22 42 cm Wireless connectivity Built in Bluetooth 4 0 BLE Power Supply Charge input Micro USB DC5V Battery type Rechargeable lithium polymer battery Charge time 2 hours Blood Pressure Technology Oscillometric method Pressure measurement range 0 300 mmHg Pressure measurement accuracy 3mmHg or 2...

Page 13: ...Network topology Point to Point Band width 1Mbps Durable period Expected service life 3 years 7 Maintenance and Cleaning 7 1 Maintenance To protect your product from damage please observe the following Store the product and the components in a clean safe location Do not wash the product and any components or immerse them in water Do not disassemble or attempt to repair the product or components Do...

Page 14: ...erence and 2 this device must accept any interference received including interference that may cause undesired operation Note This equipment has been tested and found to comply with the limits for a Class B digital device pursuant to part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference in a residential installation This equipment generate...

Page 15: ...on equipment may affect the performance of this product Other products that have RF transmitter or source may affect this product e g cell phones PDAs and PCs with wireless function Guidance and Declaration Electromagnetic Emissions The Blood Pressure Monitor is intended for use in the electromagnetic environment specified below The customer or the user of the Blood Pressure Monitor should assure ...

Page 16: ...ions on power supply input lines IEC 61000 4 11 0 UT 100 dip in UT for 0 5 cycle at 0 45 90 135 180 225 270 and 315 0 UT 100 dip in UT for 1 cycle at 0 70 UT 30 dip in UT for 25 30 cycles at 0 0 UT 100 dip in UT for 250 300 cycles at 0 N A Power frequency 50 60 HZ magnetic field IEC 61000 4 8 30 A m 50 60Hz 30 A m 50 60Hz Power frequency magnetic fields should be at levels characteristic of a typi...

Page 17: ...res objects and people a The ISM industrial scientific and medical bands between 0 15 MHz and 80 MHz are 6 765 MHz to 6 795 MHz 13 553 MHz to 13 567 MHz 26 957 MHz to 27 283 MHz and 40 66 MHz to 40 70 MHz The amateur radio bands between 0 15 MHz and 80 MHz are 1 8 MHz to 2 0 MHz 3 5 MHz to 4 0 MHz 5 3 MHz to 5 4 MHz 7 MHz to 7 3 MHz 10 1 MHz to 10 15 MHz 14 MHz to 14 2 MHz 18 07 MHz to 18 17 MHz 2...

Page 18: ... maintaining a minimum distance between portable and mobile RF communications equipment transmitters and the Blood Pressure Monitor as recommended below according to the maximum output power of the communications equipment Rated max output power of transmitter W Separation distance according to frequency of the transmitter m 150kHz to 80MHz P V d 5 3 1 80MHz to 800MHz P E d 5 3 1 800MHz to 2 7GHz ...

Page 19: ...0 Version A Oct 2019 Shenzhen Viatom Technology Co Ltd 4E Building 3 Tingwei Industrial Park No 6 Liufang Road Block 67 Xin an Street Baoan District Shenzhen 518101 Guangdong China www viatomtech com info viatomtech com ...

Reviews: