Guidance and manufacturer’s declaration – electromagnetic immunity – for all
EQUIPMENT and SYSTEMS
Guidance and manufacturer´s declaration – electromagnetic immunity
The Model Insight Infrared Video Goggles are intended for use in the electromagnetic
environment specified below. The customer or the user of the Model Insight Infrared Video
Goggles should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic
environment –
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15
kV air
± 8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15
kV air
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30 %.
Radiated RF
Immunity IEC
61000-4-3
10 V/m
10 V/m
N/A
Electrostatic
transient / burst
IEC 61000-4-4
± 2 kV for power
supply lines
100 kHz repetition
frequency
± 1 kV for input/
output lines
N/A
N/A
Surge
IEC 61000-4-5
± 0.5 kV, ± 1 kV
differential mode
line-line
N/A
N/A
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
0 % UT
(100 % dip in UT )
for 0.5 cycle at 0°, 45°,
90°, 135°,180°, 225°,
270°, and 315°
0 % UT
(100 % dip in UT )
for 1 cycle at 0°
70 % UT
(30 % dip in UT )
for 25/30 cycles at 0°
N/A
N/A
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m, 50/60Hz
30 A/m, 50/60Hz
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment.
Note: UT is the a. c. mains voltage prior to application of the test level.