Care and Maintenance
NeoGraft 2.0® User Manual
7-23
7.9. Replacing the Extraction
Handpiece Contra-Angle Head
Nozzle
The nozzle is located on the tip of the contra-angle head, as shown
in
7-1. The nozzle should be replaced if a vacuum leak is
detected from the contra-angle head.
Figure 7-1. Contra-Angle Head Nozzle
Caution
The screwdriver, provided for performing the nozzle
replacement, is not sterile and cannot be sterilized.
Make sure to clean, lubricate, and sterilize the contra-angle head
after replacing the nozzle.
To replace the nozzle of the extraction handpiece contra-angle
head:
1.
Verify that the contra-angle head is disconnected and removed
from the handpiece body.
2.
Insert the screwdriver into the two slots of the contra-angle
head nozzle.