Care and Maintenance
NeoGraft 2.0® User Manual
7-3
6.
Soak the waste jar and all of its components in an ultrasonic
bath, filled with ANIOS CLEAN EXCEL D 0.5% enzymatic
detergent solution (25 ml for 5 lit of water) at room
temperature
(15C - 25°C) for a minimum of 10 minutes.
7.
Remove the components from the ultrasonic bath and clean
each component with an appropriate brush for a minimum of
20 seconds:
•
Use a bottle brush to brush inside the glass jar.
•
Use an interdental brush to brush inside the cylindrical
parts of the black stopper.
•
Use a toothbrush to brush inside and outside of all of the
other parts.
8.
Rinse each part with lukewarm running tap water WFI (Water
for injection) for a minimum of 10 seconds.
9.
Dry each part, using a disposable wipe.
10.
Visually inspect the parts and make sure that they are clean,
intact and in good condition.
•
Any soiled part must be re-cleaned.
•
Any damaged (e.g. broken, twisted, rusty) part should be
discarded.