Care and Maintenance
NeoGraft 2.0® User Manual
7-21
Note
Quality of sterilization is dependent on the
cleanliness of the components. Make sure to
perfectly clean each item befor performing the
sterilization.
To sterilize each component:
1.
Pack each item in a pouch suitable for steam sterilization.
2.
Perform autoclave sterilization, using the following settings:
•
Temperature: 135°C (273.2°F)
•
Duration: 3 minutes
•
Dry time: 20 minutes