Troubleshooting - daily use
Problem
Checklist
Is the light flashing on the hand control?
Reset the system by simultaneously pressing the
and
buttons
on
the hand control. You will hear 10 rhythmic beeps. Press and hold the but-
tons until the beeping stops.
VENDLET Standard does not work
Is the control unit connected to 230V outlet and is the power on?
Is the green diode on the control unit illuminated?
Are all plugs properly connected on the control unit?
Is the green diode on the hand control illuminated?
Double-click the
button to activate the system.
One of the bars does not work
Are all plugs properly connected on the control unit?
Is the cable to the bar intact?
One or both bars do(es) not turn when the client is lying in bed
Is the turning sheet running under the bars?
The sheet must be over the bars.
Are all plugs properly connected on the control unit?
The hand control buttons are not working
Are all plugs properly connected on the control unit?
Double-click the
button to activate the system.
Is the correct hand control used? The VENDLET Standard only works with
the VENDLET Standard hand control (see page 8).
One of the bars will only turn in one direction
Are all plugs properly connected on the control unit?
Is the cable to the bar intact?
The symbols on the hand control do not match the action when VENDLET
Standard is activated
Are all plugs properly connected on the control unit?
The actuators cannot be raised/ lowered or do not move together
Are all plugs properly connected on the control unit?
Contact a person with relevant training, your distributor or
Vendlet ApS if the above does not solve the problem.