PART 3 - Technical manual
27
Maximum loads
Mounting
Grab the motor bar when lifting VENDLET Standard - not the
support bars.
Attaching a side
The sides are determined by the supine position of the client.
The red bar must be placed on the right side and the blue bar
on the left side of the client.
The bars must be in upper position during mounting.
1. Attach the mounting plates at the top with a button head
bolt. Do that on both ends.
2. Place the side on the steel frame of the bed.
2.
In home care situations the weight of other equipment is esti-
mated to be 25 kg. In this case the SWL of the bed must be at
least 317 kg to be able to handle a user weight of 250 kg:
Max. user
weight
VENDLET
Standard
Other
equipment
Required
Bed SWL
250 kg
+
42 kg
+
25 kg
=
317 kg
In intensive care situations the weight of other equipment is
estimated to be 55 kg. In this case the SWL of the bed must be
at least 347 kg to be able to handle a user weight of 250 kg:
Max. user
weight
VENDLET
Standard
Other
equipment
Required
Bed SWL
250 kg
+
42kg
+
55 kg
=
347 kg
1.
The max. user weight for VENDLET Standard is
250 kg.
Since VENDLET Standard is always mounted on
beds, it is important to consider the whole solution.
You should always check the safe working load
(SWL) of the bed and ensure that the combined
weight of the client, VENDLET Standard and any
other equipment on the bed never exceeds this
limit.