background image

1.  Introduction 

 
The  heating  plate  type  RC  is  designed  for  general  use,  for  all  those  applications  that  require  the  heating  of  samples  in 
specific containers positioned on the heating plate of the instrument. 
The temperature of the heating plate is controlled by an analogical bulb thermoregulator fitted directly into the heating plate. 
This makes it possible to obtain a constant control of the heating plate temperature from 50 to 370 °C. 
 

NOTE

: The container of the product used must be compatible with the temperature used. Normally, in these cases, Pyrex 

glass containers are used. The use of the heating plate with high temperatures may cause variations of the superficial colour 
but this doesn’t change the characteristics of thermal, mechanical and chemical resistance. 

2.  Assembly and installation 

 
Check the integrity of the unit after unpacking. The box includes: 

 

RC Heating Plate 

 

Power supply cord 

 

Instruction manual 

2.1 Electrical connections 

After having unpacked the instrument, place the unit on the laboratory bench.  
Before connecting the instrument to the power supply, make sure that the values on the rating plate correspond to those of 
the power supply. Connect the unit to the power supply using the transformer supplied. 
Ensure that the socket and the relative cut-off device conform to current safety norms and are easy to reach. 

2.2 Start-up 

Rotate the temperature knob completely to the left. Then, set the temperature by turning the dedicated knob. 

3.  Operating controls 

 

TEMPERATURE KNOB 

To select the temperature of the heating plate turn the corresponding knob (Heating). 
The temperature can be programmed between 50 and 370 °C.

 

ON-OFF SWITCH 

The on-

off switch turns the unit on and off. If the switch is in the “OFF” position the 

unit is off; if the switch is in the “ON” position the unit is on. 
Always turn the unit off after use. 

3.1 Error message 

The unit is fitted with safety devices which cut-off the power supply to the heating plate in the case of malfunctions. 
 

Display    

Cause 

Flashing Heating LED (once every second) 

Overheated heating plate or thermocouple open 

Flashing Heating LED (2 times every second) 

Fault in the temperature reading circuit 

Flashing Heating LED (once every 3 seconds) 

Thermocouple wires reversed or white wire of 
potentiometer board not connected 

 
Should any of the above occur, please contact your nearest VELP Scientifica service centre. 

4.  Maintenance 

 

No routine or extraordinary maintenance is necessary apart from periodically cleaning the unit as described in this manual. 
In compliance with the product guarantee law, replace the fuses and repairs to our units must be carried out in our factory, 
unless previously agreed otherwise with local distributors and using Velp spare part only. 
The instrument must be transported in its original packaging and any indications present on the original packaging must be 
followed (e.g. palletized). 
It  is  the  responsibility  of  the  user,  to  properly  decontaminate  the  unit  in  case  of  hazardous  substances  remaining  on  the 
surface  or  interior  of  the  device.  If  in  doubt  about  the  compatibility  of  a  cleaning  or  decontamination  product,  contact  the 
manufacturer or distributor. 

4.1 Cleaning 

Disconnect the unit from the power supply and use a cloth dampened with an non-inflammable non-aggressive detergent. 
 
 
 
 
 
 
 

EN 

Summary of Contents for RC F20700431

Page 1: ...uto urbano secondo quanto previsto dalla Direttiva 2002 96 CE Ne pas recycler l appareil comme déchet solide urbain conformément à la Directive 2002 96 CE No tirar el aparato en los desechos urbanos como exige la Directiva 2002 96 CE Dieses Gerät unterliegt der Richtlinie 2002 96 EG und darf nicht mit dem normalen Hausmüll entsorgt werden This unit must be used for laboratory applications only The...

Page 2: ...efactora hasta 370 C Temperaturbereich Heizplatte bis zu 370 C The heated solution may release toxic dangerous or poisonous gases Adequate safety measures must be taken in accordance with the safety regulations in force including the presence of hood and personal protective equipment masks gloves goggles etc Le sostanze riscaldate potrebbero emanare gas tossici e o pericolosi e o velenosi Adeguate...

Page 3: ...sbuchse und mit einem Schraubendreher heben Sie die kleine Abdeckung auf dem Sicherungskasten It is responsibility of the user appropriately decontaminate the instrument in case of dangerous substances fall on or in it It is also responsibility of the user to use safety substances for cleaning or decontaminating which do not react with internal parts of the instrument or with the material containe...

Page 4: ...ECHANGE 10 1 INTRODUCCIÓN 11 2 MONTAJE E INSTALACIÓN 11 2 1 CONEXIÓN A RED ELÉCTRICA 11 2 2 ENCENDIDO 11 3 CONTROLES DE FUNCIONAMIENTO 11 3 1 MENSAJE DE ERROR 11 4 MANTENIMIENTO 11 4 1 LIMPIEZA 11 5 CARACTERÍSTICAS TÉCNICAS 12 6 ACCESORIOS REFACCIONES 12 1 EINFÜHRUNG 13 2 MONTAGE UND INSTALLATION 13 2 1 ANSCHLUSS AN DAS STROMNETZ 13 2 2 INBETRIEBNAHME 13 3 BEDIENUNGSELEMENTE 13 3 1 FEHLERMELDUNGEN...

Page 5: ...ing plate turn the corresponding knob Heating The temperature can be programmed between 50 and 370 C ON OFF SWITCH The on off switch turns the unit on and off If the switch is in the OFF position the unit is off if the switch is in the ON position the unit is on Always turn the unit off after use 3 1 Error message The unit is fitted with safety devices which cut off the power supply to the heating...

Page 6: ...d A00000351 Handle for AluBlock removal A00000382 Extension for support rod ARE AREX A00000375 Hemispheric bowl 5L flasks plate 135 A00000374 Hemispheric bowl 3L flasks plate 135 A00000373 Hemispheric bowl 25ml flasks plate 135 A00000346 PTFE Safety cover for bowl 1000 ml A00000345 PTFE Safety cover for bowl 500 ml A00000344 PTFE Safety cover for bowl 250 ml A00000343 PTFE Safety cover for bowl 10...

Page 7: ... Posizione OFF lo strumento è spento se l interruttore è posto su posizione ON lo strumento è acceso L interruttore generale consente di scollegare completamente lo strumento dalla rete di alimentazione quando lo strumento non viene utilizzato al fine di ridurre gli sprechi di energia 3 1 Messaggio di errore L unità è dotata di protezioni interne che interrompono l alimentazione elettrica alla pia...

Page 8: ...icambio A00001069 Asta di sostegno A00000351 Maniglia per rimozione AluBlock A00000382 Estens per asta di sostegno ARE AREX A00000375 Calotta sferica palloni 5L testa 135 A00000374 Calotta sferica palloni 3L testa 135 A00000373 Calotta sferica palloni 25ml testa 135 A00000346 Copertura PTFE calotta sferica 1000 ml A00000345 Copertura PTFE calotta sferica 500 ml A00000344 Copertura PTFE calotta sfe...

Page 9: ...énéral permet de mettre complètement l instrument hors circuit quand l instrument n est pas utilisé afin d économiser de l énergie électrique 3 1 Le message d erreur L appareil est équipé de dispositifs de sécurité qui coupure de l alimentation de la plaque chauffante dans le cas d un mauvais fonctionnement Afficher La cause LED chauffage clignotante une fois par seconde Plaque chauffante surchauf...

Page 10: ...de rechange A00001069 Support aux enchères A00000351 Poignée pour l enlèvement d Alublock A00000382 Rallonge pour tige de support ARE AREX A00000375 Bouchon sphérique 5L tête 135 A00000374 Bouchon sphérique 3L tête 135 A00000373 Bouchon sphérique 25ml tête 135 A00000346 Couvercle sphérique en PTFE 1000 ml A00000345 Couvercle sphérique en PTFE 500 ml A00000344 Couvercle sphérique en PTFE 250 ml A00...

Page 11: ...e energía eléctrica 3 1 Mensaje de error El aparato está equipado con dispositivos de seguridad que corta la fuente de alimentación a la placa de calefacción en el caso de un mal funcionamiento Pantalla Cause Parpadeo LED calefacción un parpadeo por segundo Sobrecalentamiento de la placa de calentamiento o termopar abierto Parpadeo LED calefacción dos parpadeos por segundo Fallo en el Sistema de l...

Page 12: ...0001069 Barra soporte A00000351 Mango para la eliminación de AluBlock A00000382 Ext para varilla de soporte ARE AREX A00000375 Bolas esféricas de 5L placa135 A00000374 Bolas esféricas de 3L placa 135 A00000373 Bolas esféricas de 25ml placa 135 A00000346 Cubierta PTFE casquete esférico 1000 ml A00000345 Cubierta PTFE casquete esférico 500 ml A00000344 Cubierta PTFE casquete esférico 250 ml A0000034...

Page 13: ...eschaltet Schalten Sie das Gerät nach Gebrauch stets aus um Energie zu sparen 3 1 Fehlermeldungen Das Gerät ist mit Sicherheitseinrichtungen die die Stromversorgung der Heizplatte im Falle von Störungen cut off ausgestattet Displayanzeige Fehlfunktion Blinkende Heizungs LED einmal pro Sekunde Blinkende Heizungs LED einmal pro Sekunde Blinkende Heizungs LED 2 mal jede Sekunde Fehler im Temperaturme...

Page 14: ...1069 Stativstab A00000351 Handle for AluBlock removal A00000382 Extension for support rod ARE AREX A00000375 Hemispheric bowl 5L platte 135 A00000374 Hemispheric bowl 3L platte 135 A00000373 Hemispheric bowl 25ml platte 135 A00000346 PTFE Sicherheitsabdeckung für 1000 ml A00000345 PTFE Sicherheitsabdeckung für 500 ml A00000344 PTFE Sicherheitsabdeckung für 250 ml A00000343 PTFE Sicherheitsabdeckun...

Page 15: ...re responsabilité que le produit est conforme aux normes suivantes Nosotros casa fabricante VELP Scientifica declaramos bajo nuestra responsabilidad que el producto es conforme con las siguientes normas Der Hersteller VELP Scientifica erklärt unter eigener Verantwortung dass das Gerät mit folgenden Normen übereinstimmt EN 61010 1 2010 EN 61326 1 2013 2011 65 EU RoHS 2012 19 EU RAEE EN61010 2 051 2...

Page 16: ...Analyzers Digestion Units Distillation Units Solvent Extractors Fiber Analyzers Dietary Fiber Analyzers Oxidation Stability Reactor Consumables Analytical Instruments Analizzatori Elementari Digestori e Mineralizzatori Distillatori Estrattori a Solventi Estrattori di Fibra Estrattori di Fibra Dietetica Reattore di Ossidazione Consumabili Laboratory Equipment Magnetic Stirrers Heating Magnetic Stir...

Reviews: